9.7 How to Set the Instrument Constant Value
To set the Instrument constant which is obtained in section 9.1 “Checking and Adjusting of
Instrument Constant”, follow as below.
Note:
Each of the Prism mode and Non-prism mode has instrument constant.
You must obtain the instrument constant of prism mode. If you reset the instrument constant
of the prism mode, you must vary the non-prism constant (Non-prism-short, Non-prism-long)
by the same quantity as the amount of increase or decrease of the instrument constant of
the prism mode.
Operating procedure
Operation
Display
1
Press the [F5] key from the main menu.
[F5]
2
Press [F4] (Instrument constant) key.
[F4]
3
Select constant value.
Example: Prism
[F1]
4
Press [F5] (YES) key .
[F5]
5
Input value and press [ENT] key. *1),2)
Input
value
[ENT]
6
Press [F5](OK) key.
The display returns to main menu.
[F5]
*1) Refer to Section 2.9 How to Enter Numerals and Alphabet Letters .
*2) To cancel the setting, press the [ESC] key.
Adjustment
F1 V0/Axis (Measurement)
F2 V0/Axis (Constant list)
F3 Date Time
F4 Instrument constant
↓
Instrument Constant
F1 PRISM
F2 NON-PRISM-SHORT
F3 NON-PRISM-LONG
Instrument Constant
EDM OFFSET (mm)
0.0
Modify
YES NO
Instrument Constant
EDM OFFSET (mm)
0.0
This is not Prism Offset!
EXIT BS
Instrument Constant
EDM OFFSET (mm)
1.2
The old value will be erased!
OK CANCEL