SOL-96.110.061 / 031221
TOPAZ Manganese
Troubleshooting
56
31
Water default
– Check that the analyzer is supplied
with water.
– Check if the filling valve (SVB)
switches. If this is not the case,
perform the following steps:
– Check that the cable is correctly
plugged in at the back of the
solenoid valve.
– If necessary, replace the membrane.
– Check sample tubes.
32
Pressure measure default -
Measure < 0
– Check that the analyzer is supplied
with water.
– Check if the filling valve (SVB)
switches. If this is not the case,
perform the following steps:
– Check that the cable is correctly
plugged in at the back of the
solenoid valve.
– If necessary, replace the membrane.
– Check sample tubes.
33
Optical setting Default - Info
setting not received
– Observe the “Live optical measure”
value in the <DIAGNOSTIC> tab. With
a functioning measuring card, the
value changes continuously.
– If the value does not change, restart
the analyzer and check again.
– If the value changes, launch a new
cycle and check if the error is still
displayed.
– If the error persists, contact customer
support.
34
Measure default - measure < 0
(optical measure higher than
baseline measured at beginning
of measuring cycle).
– Check if the reagents have been
stored incorrectly.
– Check that reagents are correctly
pumped into the photometer.
– Check that the stirrer turns.
35
Measure stability default
– Check that the optical fibers are well
connected to their holder on the door.
No.
Text shown on display
Corrective action