SOL-96.110.061 / 031221
TOPAZ Manganese
Troubleshooting
60
8.3. Saving Diagnostic Data to USB Stick
The TOPAZ Ammonium permanently records measurement and
diagnostic data and stores it in a non-volatile memory. If required,
this data can be transferred to a USB stick.
Transfer to
USB stick
To transfer the measurement and diagnostic data to a USB stick,
proceed as follows:
1
Open the analyzer door.
2
Insert the USB stick into the USB port [A] on the back of the
panel PC.
3
Open the <TOOLS> tab and press the <Log files copy> button.
Enter the password “0712”, if necessary.
4
On the next screen, press OK to start data transfer.
Contents of the
USB stick
After the data transfer, you will find a folder named "CRISTAL_LOG"
on the USB stick, which contains the following subfolders:
CristalErrorPrgLog: Alarms and errors.
CristalLog: Measured values with date, time, channel number
and diagnostic data. In case of support requests, please have
these files ready.
UserLogFileDir: Measured values with date, time and channel
number
A
USB port