SOL-96.110.061 / 031221
TOPAZ Manganese
Troubleshooting
53
10
Error calibration cycle execution – Observe the “Live optical measure”
value in the <DIAGNOSTIC> tab. With
a functioning measuring card, the
value changes continuously.
– If the value does not change, restart
the analyzer and check again.
– If the error persists, contact customer
support.
11
Error reading manual cycle
configuration
– Reset analyzer configuration to factory
defaults by pressing the <Recover>
button in the <+> tab.
12
Error writing manual cycle
configuration
– Reset analyzer configuration to factory
defaults by pressing the <Recover>
button in the <+> tab.
13
Error cycle execution
– Observe the “Live optical measure”
value in the <DIAGNOSTIC> tab. With
a functioning measuring card, the
value changes continuously.
– If the value does not change, restart
the analyzer and check again.
– If the error persists, contact customer
support.
14
Error reading socket
configuration
– Reset analyzer configuration to factory
defaults by pressing the <Recover>
button in the <+> tab.
15
Error writing socket
configuration
– Reset analyzer configuration to factory
defaults by pressing the <Recover>
button in the <+> tab.
16
Error socket execution
– Restart the analyzer.
17
Error reading application
configuration
– Reset analyzer configuration to factory
defaults by pressing the <Recover>
button in the <+> tab.
18
Error writing application
configuration
– Reset analyzer configuration to factory
defaults by pressing the <Recover>
button in the <+> tab.
No.
Text shown on display
Corrective action