SOL-96.110.061 / 031221
TOPAZ Manganese
Troubleshooting
52
8.2. List of Errors and Alarms
No.
Text shown on display
Corrective action
1
Unknown parameter or variable
– Contact customer support.
2
Error reading cycle configuration – Reset analyzer configuration to factory
defaults by pressing the <Recover>
button in the <+> tab.
3
Error writing cycle configuration
– Reset analyzer configuration to factory
defaults by pressing the <Recover>
button in the <+> tab.
4
Error cycle execution
– Observe the “Live optical measure”
value in the <DIAGNOSTIC> tab. With
a functioning measuring card, the
value changes continuously.
– If the value does not change, restart
the analyzer and check again.
– If the error persists, contact customer
support.
5
Error reading cycle zero
configuration
– Reset analyzer configuration to factory
defaults by pressing the <Recover>
button in the <+> tab.
6
Error writing cycle zero
configuration
– Reset analyzer configuration to factory
defaults by pressing the <Recover>
button in the <+> tab.
7
Error cycle zero execution
– Observe the “Live optical measure”
value in the <DIAGNOSTIC> tab. With
a functioning measuring card, the
value changes continuously.
– If the value does not change, restart
the analyzer and check again.
– If the error persists, contact customer
support.
8
Error reading calibration cycle
configuration
– Reset analyzer configuration to factory
defaults by pressing the <Recover>
button in the <+> tab.
9
Error writing calibration cycle
configuration
– Reset analyzer configuration to factory
defaults by pressing the <Recover>
button in the <+> tab.