SOL-96.110.061 / 031221
TOPAZ Manganese
Maintenance
35
Reagent
consumption
LR variant
At the standard measurement interval, the following amounts of
reagents are consumed:
Reagent
consumption
HR variant
At the standard measurement interval, the following amounts of
reagents are consumed:
Replacing
reagents
1
Prepare the reagents according to the separate method sheet.
2
Insert the reagent tubes all the way to the bottom of the corre
-
sponding bottles (see
).
3
Check that there is an additional small hole in the lid of each
bottle that allows air to enter.
Final steps
4
Prime the reagent tubes (see
Priming the Reagent Tubes, p. 25
).
5
Perform a manual measurement.
If the result is satisfactory, no calibration is needed.
6
If necessary, calibrate the instrument (see
).
Reagent
Consumption
Hydroxylamine Chlorhydrate
100 g/L + 1.0 mg/L Mn
3.2 L/month
Triethanolamine 20%
3.2 L/month
Buffer ammoniacal pH 9.2
3.2 L/month
H
2
SO
4
2N
3.2 L/month
Potassium cyanide
3.2 L/month
Pan + Triton X-100
1 L/month
Reagent
Consumption
Hydroxylamine Chlorhydrate
100 g/L
3.2 L/month
Triethanolamine 20%
3.2 L/month
Buffer ammoniacal pH 9.2
3.2 L/month
H
2
SO
4
2N
3.2 L/month
Potassium cyanide
3.2 L/month
Pan + Triton X-100
3.2 L/month