SOL-96.110.061 / 031221
TOPAZ Manganese
Maintenance
40
6.6. Manual Measurement and Calibration
Methods of
calibration
The TOPAZ Manganese supports two methods of calibration: A
standard calibration or an online calibration. Seres OL recommends
to use the standard calibration.
If you want to use the online calibration, contact customer service for
details.
Entering the
maintenance
menu
1
To stop the current measuring cycle, press the button on the
main screen and enter the password “0712”. In the next dialog,
confirm with “OK”.
2
From the main screen, open the <TOOLS> tab and press the
<Maintenance>
button.
The <MAINTENANCE> menu shows the following entries:
MEASURE
DIAGNOSTIC
+
TOOLS
Maintenance
Electrical test
Maintenance
Maintenance
Electrical test
MAINTENANCE
DIAGNOSTIC
Stopped
Zero calibration
Calibration
Calibration in line
Manual Measure