Troubleshooting
54
22
Error execution on IO card
– Check if all I/O boards work properly:
– Navigate to <TOOLS>/
<Electrical test>/
– Open the <RELAYS> tab and test
the relays.
– Open the <INPUT> tab and test the
digital inputs.
– If the error persists, contact customer
support.
23
Error reading JBUS slave card
configuration
– Reset analyzer configuration to factory
defaults by pressing the <Recover>
button in the <+> tab.
24
Error writing JBUS slave card
configuration
– Reset analyzer configuration to factory
defaults by pressing the <Recover>
button in the <+> tab.
25
Error reading on JBUS slave
card
– Check request compliance with JBUS
protocol.
26
Error treatment on JBUS slave
card
– Check request compliance with JBUS
protocol.
27
Error reading measuring card
configuration
– Reset analyzer configuration to factory
defaults by pressing the <Recover>
button in the <+> tab.
28
Error reading measuring card
configuration
– Reset analyzer configuration to factory
defaults by pressing the <Recover>
button in the <+> tab.
29
Temperature setpoint not
reached
– Check that the ambient and/or the
sample temperature does not exceed
the temperature setpoint.
30
Vessel temperature unstable
– Check that the ambient and/or the
sample temperature does not exceed
the temperature setpoint.
No.
Text shown on display
Corrective action