SOL-96.110.081 / 260422
TOPAZ Iron
Troubleshooting
53
19
Error application execution
– Observe the “Live optical measure”
value in the <DIAGNOSTIC> tab. With
a functioning measuring card, the
value changes continuously.
– If the value does not change, restart
the analyzer and check again.
– Check if all I/O boards work properly:
– Navigate to <TOOLS>/
<Electrical test>/
– Open the <RELAYS> tab and test
the relays.
– Open the <INPUT> tab and test the
digital inputs.
– If the error persists, contact customer
support.
20
Error reading on IO card
– Check if all I/O boards work properly:
– Navigate to <TOOLS>/
<Electrical test>/
– Open the <RELAYS> tab and test
the relays.
– Open the <INPUT> tab and test the
digital inputs.
– If the error persists, contact customer
support.
21
Error writing on IO card
– Check if all I/O boards work properly:
– Navigate to <TOOLS>/
<Electrical test>/
– Open the <RELAYS> tab and test
the relays.
– Open the <INPUT> tab and test the
digital inputs.
– If the error persists, contact customer
support.
No.
Text shown on display
Corrective action