SOL-96.110.081 / 260422
TOPAZ Iron
Startup
27
Check
stirrer
To check the proper function of the stirrer, perform the following
steps:
10
Open the <ACTUATOR> tab.
11
Set “SV Emptying” to <ON> and “SV Filling” to <OFF> and wait
until the instrument is empty.
12
Remove the lid from the photometer.
13
Check that the stirrer bar is turning.
14
Reinstall the lid on the photometer.
ON
ON
ON
ON
OFF
OFF
OFF
OFF
Stirrer
SV Emptying
Heating
SV Filling
ACTUATOR
4-20mA OUTPUT
PUMPS
RELAYS
INPUTS
Vessel temperature :