TOPAZ Iron
Startup
28
4.4. Run-in Period
1
Go back to the main screen.
2
Start measurement by pressing the button.
3
Let the analyzer perform one or two measuring cycles before
proceeding with the next steps.
4.5. Manual Measurement
1
Perform a manual measurement with a standard solution.
If the result of the manual measurement is satisfactory, no
calibration is necessary.
2
Calibrate the analyzer if required.
For detailed descriptions, see
Manual Measurement and Calibration,
.
4.6. Programming
Program the following parameters, if applicable:
Concentration values for thresholds 1 and 2
Definition as “high/high”, “low/high” or “low/low” thresholds,
Number of measuring cycles per channel
Measuring interval
If digital inputs are used or if the analyzer is to be controlled via
JBUS, select the appropriate control mode
JBUS slave ID (set to a number higher than 16)
For detailed descriptions of all configuration options see
Configuration of the Analyzer, p. 43
.