SOL-96.110.081 / 260422
TOPAZ Iron
Maintenance
41
6.6.2
Zero Calibration
Note:
• On multi-channel instruments, only calibrate sample stream 1.
Procedure
The procedure is similar to the manual measurement procedure
. Use deionized water in
-
stead of standard.
Results
screen
At the end of the zero calibration, a window displays the previous
(actual) and the new (calculated) offset. Confirm by pressing the
button.
6.6.3
Standard Calibration
Note:
• On multi-channel instruments, only calibrate sample stream 1.
Procedure
The procedure is similar to the manual measurement procedure
.
Begin calibration by entering the reference value of the standard
used. Pay attention to the correct concentration unit (ppm by de
-
fault). Then follow the instructions on the screen.
Results
screen
At the end of the standard calibration, a window displays the previ
-
ous (actual) and the new (calculated) calibration coefficient and the
concentration measurement readings before and after calibration.
Confirm by pressing the button.
CALIBRATION ZERO RESULTS
Actual offset
Offset Calculated
-0.1
0.2
CALIBRATION RESULTS
Actual coefficient
Coef. Calculated
Measure before calib
Measure after calib
1.040
1.047
20.15
20.00