Chapter 2 Precautions on Handling
This chapter describes the precautions to be followed in the handling of this tester. When using the
tester, take utmost care to ensure safety.
!
WARNING: The tester derives a 5 KVAC or 6KVDC test high voltage, incorrect or wrong
operation can cause accidents which will result in human injury or death. When operating the tester,
be extremely careful and observe the cautions, warnings, and other instructions given in this chapter,
keep them in mind to avoid accidents.
2.1
Operation Standards
1)
Wearing Insulation Gloves
When handling the tester, be sure to wear insulation gloves in order to protect yourself against
high voltages, even though, please do not touch the live conductor by hand when high voltage test
is conducting.
2)
Grounding
There is a protective grounding terminal on the rear panel of the tester chassis,
please use a dedicated tool to connect it to an electrical ground (safety ground,
earth). If there is no good grounding, when the circuit of the power is shorted to
the ground or any device’s power line is shorted to the ground, the tester chassis is
charged to an excessively high voltage, resulting in extreme danger, anyone who
touches such tester chassis, will result in electric shock accident, therefore, please
make sure that the protective grounding terminal is connected to the safety
ground.
3)
Test Line Connection
Press
【
STOP
】
first, confirm that DANGER lamp is not lit, then connect the test line. When
connecting the test lead wire to a DUT, please connect the DUT with the test lead wire of HV2
or RTN/LOW first. It is very dangerous if the test line connection of HV2 or RTN/Low is
incomplete or dropped, because the entire DUT will be charged with high voltage.
After high voltage plug of the high voltage test lead is inserted into HV1 and HV2, it must be
rotated 90 degrees clockwise to prevent the test lead wire from falling off.
Before testing, it is a must to check whether the test lead wire of HV1 and HV2 or RTN/LOW are
connected, loose or fall off.
4)
Stop (Pause) Test
To change the test conditions, please press the STOP once first, this will take the tester out of
the test preparation state and ensure that the DANGER lamp is off. If need to take a break or
leave the test location, please switch off the power, to prevent the accidental contact with the
start switch and cause a safety hazard.
5)
Remote Control
Be very careful when operating the tester in remote control mode, because the start and stop of
Safe ground