65
Command message
:
FUNC:SOUR:STEP <sn>:IR:RANG<range>
Query message
:
FUNC:SOUR:STEP <sn>:IR:RANG?
--
Data<range>:
Data format: integer
Data range: 0~6 (0 is AUTO, 1 is 10mA, 2 is 3mA, 3 is 300uA, 4 is 30uA, 5 is 3uA, 6 is
300nA)
--
Example:
Set the range for IR test in STEP1 to10mA.
Command message: FUNC:SOUR:STEP 1:IR:RANG 1
Query message: FUNC:SOUR:STEP 1:IR:RANG?
Return value: 1
5.2.3.6 Commands for PA Setup Function
FUNC:SOURce:STEP:PA:MESSAge
To set /inquiry about the PA message.
--
Syntax:
Command message: FUNC:SOUR:STEP <sn>:PA:MESSAge <message >
Query message: FUNC:SOUR:STEP <sn>:PA:MESSAge?
--
Data<message>:
Data format: characters
--
Example:
Set the PA message in STEP1 to HELLO!
Command message: FUNC:SOUR:STEP 1:PA:MESSAge HELLO!
Query message: FUNC:SOUR:STEP 1:PA:MESSAge?
Return value: HELLO!
FUNC:SOURce:STEP:PA:TIME
To set /inquiry about the PA state continue time.
–
Syntax:
Command message: FUNC:SOUR:STEP <sn>:PA:TIME<time>
Query message: FUNC:SOUR:STEP <sn>:PA:TIME?
--
Data<time>:
Data format: float
Data range: 0, 0.3
~
999.0s (0 is continue)
Data accuracy: 0.1s
Data unit: s
--
Example
:
Set the PA continue time in STEP1 to1s.
Command message
:
FUNC:SOUR:STEP 1: PA:TIME 1
Query message
:
FUNC:SOUR:STEP 1: PA:TIME?
Return value
:
1.0
5.2.3.7 Commands for OSC Setup Function
FUNC:SOURce:STEP:OS:OPEN
To set /inquiry about the OPEN rate for OS test.