Syntax:
Command message:
FUNC:SOUR:STEP <sn>:OS:OPEN<rate>
Query message:
FUNC:SOUR:STEP <sn>:OS:OPEN?
--
Data<rate>:
Data format: integer
Data range: 10~100
Data accuracy: 10
Data unit:
--
Example:
Set the OPEN rate for OS test in STEP1 as 50%.
Command message: FUNC:SOUR:STEP 1:OS:OPEN 50
Query message: FUNC:SOUR:STEP 1:OS:OPEN?
Return value: 50
FUNC:SOURce:STEP:OS:SHOT
To set /inquiry about the SHOT rate for OS test.
--
Syntax:
Command message:
FUNC:SOUR:STEP <sn>:OS:SHOT<rate>
Query message:
FUNC:SOUR:STEP <sn>:OS:SHOT?
--
Data< rate>:
Data format: integer
Data range: 0, 100~500 (0 is OFF)
Data accuracy: 10
Data unit:
--
Example:
Set the SHOT rate for OS test in STEP1 as 100%.
Command message: FUNC:SOUR:STEP 1:OS:SHOT 100
Query message: FUNC:SOUR:STEP 1:OS:SHOT?
Return value: 100
FUNC:SOURce:STEP:OS:GET
Outputs the capacitance.
--
Syntax:
Command message:
FUNC:SOUR:STEP <sn>:OS:
GET
The instrument will automatically sample the standard capacitance value after impedance
transformation of the current.
FUNC:SOURce:STEP:OS:STAND
To set /inquiry about the STANDARD capacitance for OS
test.
--
Syntax:
Command message: FUNC:SOUR:STEP <sn>:OS:STAND<standard capacitance >