5.2.3.3 AC Setup Fuction Commands
FUNC:SOURce:STEP:AC:VOLT
To set / inquiry about the voltage for ACW test.
--
Syntax:
Command message:
FUNC:SOUR:STEP <sn>:AC:VOLT<voltage>
Query message:
FUNC:SOUR:STEP <sn>:AC:VOLT?
--
Data<sn>:
Data format: integer
Data range: 1~50
Data accuracy: 1
--
Data<voltage>:
Data format: integer
Data range: 50~5000
Data accuracy: 1
Data unit: V
--
Example:
Set the voltage for ACW test in STEP1 as 1000V.
Command message: FUNC:SOUR:STEP 1:AC:VOLT 1000
Query message: FUNC:SOUR:STEP 1:AC:VOLT?
Return value: 1000
FUNC:SOURce:STEP:AC:UPPC
To set/inquiry about the UPPER current for ACW test.
--
Syntax:
Command message:
FUNC:SOUR:STEP <sn>:AC:UPPC<current>
Query message:
FUNC:SOUR:STEP <sn>:AC:UPPC?
--
Data<voltage>:
Data format: float
Data range: 0.001
~
120.000mA (voltage <4000V)
0.001
~
100.000mA (voltage>4000V)
Data accuracy: 0.001mA
Data unit: mA
--
Example:
Set the current for ACW test in STEP1 as 1mA.
Command message: FUNC:SOUR:STEP 1:AC:UPPC 1
Query message: FUNC:SOUR:STEP 1:AC:UPPC?
Return message: 1.000
FUNC:SOURce:STEP:AC:LOWC
To set /inquiry about the LOWER current for ACW test.
--
Syntax: