![Tonghui Electronics TH9110 Скачать руководство пользователя страница 19](http://html1.mh-extra.com/html/tonghui-electronics/th9110/th9110_operation-manual_1126462019.webp)
◇
19
Figure 4-3
4.2.1.4 Auto Range
When the auto range, that is, the withstand voltage auto range function is set to ON, at 0.6S before
the completion of the test, if
the measured current can be expressed by the low current range, the
current range is automatically changed to the low current range. Examples are shown in Figure
4-4.1 and Figure 4-4.2 and Figure 4-4.3.
1.
If the current range in TEST is set as high current range, 100mA in Figure 4-4.1.
Figure4-4.1
2.
But the actual current is
comparatively small, 0.4mA as shown in Figure 4-4.2.