        Command message: FUNC:SOUR:STEP 1:DC:RTIM 1 

        Query message:    FUNC:SOUR:STEP 1:DC:RTIM? 

                Return value:      1.0 

 

FUNC:SOURce:STEP:DC:FTIM             

To set /inquiry about the FALL time for DCW test.

 

-

Syntax: 

       

Command message: FUNC:SOUR:STEP    <sn>:DC:FTIM<time> 

          Query message: FUNC:SOUR:STEP    <sn>:DC:FTIM?

 

--

Data<time>: 

Data format: float 

Data range: 0~999.0s      (0 is OFF) 

Data accuracy: 0.1s 

Data unit: s 

--

Example: 

      Set the fall time for DCW test in STEP1 to1s 

      Command message: FUNC:SOUR:STEP 1:DC:FTIM 1 

      Query message: FUNC:SOUR:STEP 1:DC:FTIM? 

              Return value:      1.0 

 

FUNC:SOURce:STEP:DC:WTIM           

To set /inquiry about the wait time for DCW test.

 

--

Syntax: 

Command message: 

FUNC:SOUR:STEP    <sn>:DC:WTIM<time> 

Query message: 

FUNC:SOUR:STEP    <sn>:DC:WTIM? 

--

Data<time>: 

Data format: float 

Data range: 0~999.0s      (0 is OFF) 

Data accuracy: 0.1s 

Data unit: s 

     

--

Example: 

                  Set the wait time for DCW test in STEP1 as 1s. 

Command message: FUNC:SOUR:STEP 1:DC:WTIM 1   

Query message: FUNC:SOUR:STEP 1:DC:WTIM? 

Return message: 1.0 

 

FUNC:SOURce:STEP:DC:ARC         

To set /inquiry about the ARC upper current for DCW test.

 

--

Syntax: 

         

Command message: FUNC:SOUR:STEP    <sn>:DC:ARC<current > 

            Query message: FUNC:SOUR:STEP    <sn>:DC:ARC?

 

--

Data<current>: 

Data format: float 

Data range: 0, 1.0

10.0mA    (0 is OFF) 

Data accuracy: 0.1mA 

Содержание TH9110

Страница 1: ...1 TH9110 TH9110A AC DC Withstanding Voltage Resistance Tester V1 5 2018 07 Operation Manual Tonghui Electronic Co Ltd www tonghui com cn...

Страница 2: ...Basic operation 15 4 1 Interface structure overview 15 4 2 SYSTEM Setup 16 4 2 1 System Test Parameter Settings 16 4 2 2 System Environment Parameters Setting 21 4 2 3 System Communication Setup 23 4...

Страница 3: ...S232C Interface Instruction 45 5 1 2 GPIB Interface Instruction 51 5 2 Serial Port Commands Instruction 54 5 2 1 SCPI Commands 54 5 2 2 DISPlay Subsystem Commands 55 5 2 3 FUNCtion Subsystem Commands...

Страница 4: ...enable the device to work with various automatic testing systems that require high security and reliability The tester provides multiple test functions typical ones High Voltage Floating Output desig...

Страница 5: ...Operating Temperature 20 8 Humidity 80 RH Storage Ambient Temperature 10 55 Humidity 90 RH 1 2 3 Preheating Warm up time after power on 20 minutes 1 2 4 Precautions 1 Please do not locate and use the...

Страница 6: ...otectors of the tester may not be able to render their protective functions properly Do not turn OFF the power switch when the tester is delivering its test voltage you may do this only in case of eme...

Страница 7: ...Electromagnetic Compatibility Power Transient Sensitivity refer to requirements of GB6833 4 Conductance Sensitivity refer to requirements of GB6833 6 Radiative Interference refer to requirements of GB...

Страница 8: ...resulting in extreme danger anyone who touches such tester chassis will result in electric shock accident therefore please make sure that the protective grounding terminal is connected to the safety g...

Страница 9: ...to discharge after the output is cut off There is a danger of electric shock during discharge do not touch anything that could cause an electric shock As soon as the output is cut off the tester s dis...

Страница 10: ...he high voltage output terminal HV1 to HV2 or RTN LOW the entire housing of the tester must be reliably connected to the ground 11 Do not connect external voltage to the test terminal Do not apply a v...

Страница 11: ...ristics etc If different type of fuse is used or short circuit then the tester may be damaged 6 AC Power source The AC input power used by this tester is 100V 240VAC frequency is 47Hz 63Hz If the powe...

Страница 12: ...rnally connect to USB storage 4 Brand and Model 5 PrtScn key Screenshot key it is used to capture the current screen image to the USB memory and USB need to be inserted into the front panel jack in ad...

Страница 13: ...potential terminal of high voltage output this is the high voltage output terminal when DANGER light is on it means there is high voltage delivering touching is forbidden 13 HV2 Terminal RTN terminal...

Страница 14: ...uter 7 LAN Interface LAN communication interface can realize the communication with the computer 8 USB Device interface Through USB DEVICE communication interface the computer can control the instrume...

Страница 15: ...amme and the total steps is 1 AC means the AC withstanding voltage test interface AC parameter means that other parameters are AC withstanding voltage test parameters The third line in the interface i...

Страница 16: ...lated system settings 3 Press key to move the cursor to the parameters you want to set Change the parameter settings with F1 F6 keys or numeric keys 4 If you need to input with the numeric keys press...

Страница 17: ...an be set to continue the next test or press START to restart or press STOP first then press START to start the test PassHold 0 2 99 9S 0 5S Set the duration of the buzzer sound when in PASS StepHold...

Страница 18: ...STOP key before pressing the START key to restart the test 4 2 1 3 Hardware Software Compensation Due to the output impedance of the test signal source the load effect that is the load change the outp...

Страница 19: ...easured current can be expressed by the low current range the current range is automatically changed to the low current range Examples are shown in Figure 4 4 1 and Figure 4 4 2 and Figure 4 4 3 1 If...

Страница 20: ...ting has three options namely OFF ON and FLOAT only TH9110 has high voltage floating output function OFF and ON means the function is turned off and on FLOAT means that the two high voltage output HV1...

Страница 21: ...hus ensuring the safety of operator As shown in Figure 4 7 Figure 4 6 Figure 4 7 Relationship of HV1 HV2 Terminal to the Ground 4 2 2 System Environment Parameters Setting Operation Instructions 1 Pre...

Страница 22: ...ON or OFF LCD Bri 1 10 5 Adjust the screen brightness Fail Vol OFF ON ON Set the FAIL beep to ON or OFF Date Set Set the year month and day as needed Time Set Set 24 hour time as needed Table 4 2 Key...

Страница 23: ...be set and change the parameter setting by F1 F6 or numeric key System communication setup as shown in Figure 4 10 Figure 4 10 system communication setting As can be seen from the above figure the co...

Страница 24: ...up Test procedures current setting step number total number of test steps The change of the test steps are as shown in the table below Key Function Explanation F1 INSERT Add a new test item The curren...

Страница 25: ...imit 0 001mA Hi Limit 0 CLOSE CLOSE Set the low limit of the test current The low limit must be high limit or Close Arc Limit 1mA 20mA 0 CLOSE CLOSE Set the allowable AC arc current high limit CLOSE m...

Страница 26: ...stand voltage testing Hi Limit 0 0001mA 25mA 500 0uA Set the high limit of the test current The voltage is 1 5kV the maximum current can reach 20mA the voltage is 1 5kV the maximum current can reach 2...

Страница 27: ...age When Ramp Time is set to Close the voltage rise is completed during the first cycle of the test DwellTime 0 1 999 0s 0 CLOSE CLOSE Set the time needed for dwell the current high limit and low limi...

Страница 28: ...ltage Close the ramp time the voltage ramp is completed during the first cycle of the test MeasTime 0 3 999 0s 0 Continue 3 0S Set the measurement time for the insulation resistance test 0 indicates t...

Страница 29: ...5M 3mA 0 3 3mA 0 5M 4 5M 3 0M 15 0M 300uA 30 300uA 3 0M 15 0M 10 0M 45M 30uA 3 30uA 10 0M 45M 35 0M 450M 3uA 0 3 3uA 45M 0 45G 0 40G 4 5G 300nA 20 300nA 0 40G 4 9G 4 0G 50 0G Table 4 6 4 3 4 Pause Mod...

Страница 30: ...ng the numeric keys or you can obtain a value as a standard value by sampling Table 4 7 Sampling Move the cursor to Sampling as shown in Figure 4 17 Input the standard value by using numeric key then...

Страница 31: ...for details The upper one is the high voltage output voltage and the withstand voltage s unit is kilovolts kV The middle one is the test value the withstand voltage testing shows current unit is mill...

Страница 32: ...erface Figure 4 18 3 IR Test Interface F1 4 2 1 4 2 co m F2 co m F3 co m F4 F1 F2 co m F3 co F4 co m F5 4 2 1 4 2 co m F6 co m F1 4 2 1 F2 co m F3 co m F4 F1 F2 co m F3 co F4 co m F5 4 2 1 F6 co m F1...

Страница 33: ...arts to measure the leakage current of the test cable and display the current use the present test value as the zero value 3 After pressing F1 offset key the DANGER indicator flashes and the voltage o...

Страница 34: ...After pressing START to start the test the third column shows the present voltage test value and the fourth column shows the actual measured current value or resistance value As shown in Figure 4 20 4...

Страница 35: ...tage output 3 Finally connect the test cable of low potential RTN LOW or HV2 to the DUT first and then connect the test cable of high potential HV1 to the DUT 4 5 4 Test Procedures 4 5 4 1 AC DC IR Te...

Страница 36: ...repare for the test Press the START key to start the test when this key is pressed there will be a high voltage output and the DANGER light will flash The first line of the large font shows the actual...

Страница 37: ...s 350P 400P 88 and it is recommended to take 60 3 SHORT value the low limit is 450P 400P 112 the high limit is550P 400P 138 and it is recommended to take 125 4 6 Voltage Breakdown Test 4 6 1 Brief des...

Страница 38: ...age value Meas Mode AC DC AC Measurement mode selection AC DC Ve 0 05 5kV 0 05kV Set voltage ending value Vs 0 05kV Termination Voltage 0 05kV Set voltage starting value Ramp Time 0 999 0 CLOSE CLOSE...

Страница 39: ...File Storage Operation Instructions 1 Press FILE key to enter the file management interface as shown in Figure 4 24 2 Through the F1 key F2 key to switch between internal files and external files 3 Mo...

Страница 40: ...esult output signal Signals can be divided into two broad categories internal output signals and external input signals The internal output can be divided into test result output and contact signal ou...

Страница 41: ...stem generates an internal error 16 EOT Output When this signal is HIGH it means the test is in process when this signal is LOW it means the test is finished or in standby 17 EOS Output When this sign...

Страница 42: ...42 Take the External Power Supply as an Example...

Страница 43: ...43...

Страница 44: ...nal trigger signal EXT_STRAT time needed from start to EOT signal is cleared 20mS T3 Time setup for the trigger delay T4 Time needed for each test steps STEP T5 10mS PASS_FAIL signals stable waiting t...

Страница 45: ...d which can also be called the asynchronous serial communication standard which is used to realize data communication between a computer and a computer and between a computer and a peripheral device R...

Страница 46: ...omputer The tester is connected to the computer as shown As can be seen from the above figure the pin definition of this tester is the same as that of the 9 core connector serial interface pin used by...

Страница 47: ...do not receive the returned characters there may be factors 1 Serial port connection failure 2 Check that the RS232 port function is enabled 3 The tester is executing a bus command and is temporarily...

Страница 48: ...main function can be used by the user to expand the communication function arbitrarily while other sub functions illustrate how to use the serial port for input and output of the string define PORT 0...

Страница 49: ...PORT 256 c read_port PORT break if n break if c ps ps else printf nE10 Write Echo Error exit 1 send_port PORT n send command end symbol delay 2 while check_stat PORT 256 read_port PORT read string fr...

Страница 50: ...ar read_port int port union REGS r r x dx port serial port r h ah 2 int14 function2 read character int86 0x14 r r if r h ah 128 if ah 7 be set mean trans error printf nE01 Serial port read error exit...

Страница 51: ...board is an option The control command system is open and the user can use the computer operation interface provided by the tester or can program the system according to the control command to achieve...

Страница 52: ...52 Figure5 1 GPIB Connector Pin Structure Diagram GPIB Cable Connection Method 1 Tester Back Connector...

Страница 53: ...er contact functions T5 Basic talk function Talk only function Talking cancelled when MLA Do not support serial roll call L4 Basic listen function Listen cancelled when MTA No listen only function RL1...

Страница 54: ...3 The data of the command is an ASCII string The default format of the system is integer or floating point number The default data unit does not appear in the commands 4 The command end must have an...

Страница 55: ...e to Standby Home Page Character can query the current page Example Set the display page to measurement display Command Syntax DISP PAGE TEST Query syntax DISPlay PAGE Return format TEST Set measureme...

Страница 56: ...STEP 1 AC VOLT 1000 Query message FUNC SOUR STEP 1 AC VOLT Return value 1000 FUNC SOURce STEP AC UPPC To set inquiry about the UPPER current for ACW test Syntax Command message FUNC SOUR STEP sn AC UP...

Страница 57: ...for ACW test Syntax Command message FUNC SOUR STEP sn AC TTIM time Query message FUNC SOUR STEP sn AC TTIM Data time Data format float Data range 0 0 3 999 0s 0 is OFF Data accuracy 0 1s Data unit s...

Страница 58: ...ge FUNC SOUR STEP 1 AC FTIM 1 Query message FUNC SOUR STEP 1 AC FTIM Return message 1 0 FUNC SOURce STEP AC ARC To set inquiry about the ARC upper current for ACW test Syntax Command message FUNC SOUR...

Страница 59: ...Data sn Data format integer Data range 1 50 Data accuracy 1 Data voltage Data format float Data range 50 6000 Data accuracy 1 Data unit V Example Set the voltage for DCW test in STEP1 to 1000V Command...

Страница 60: ...alue 1 000 FUNC SOURce STEP DC TTIM To set inquiry about the TEST time for DCW test Syntax Command message FUNC SOUR STEP sn DC TTIM time Query message FUNC SOUR STEP sn DC TTIM Data time Data format...

Страница 61: ...TIM Return value 1 0 FUNC SOURce STEP DC WTIM To set inquiry about the wait time for DCW test Syntax Command message FUNC SOUR STEP sn DC WTIM time Query message FUNC SOUR STEP sn DC WTIM Data time Da...

Страница 62: ...est in STEP1 to 1mA Command message FUNC SOUR STEP 1 DC RAMPARC 1 Query message FUNC SOUR STEP 1 DC RAMPARC Return message 1 0 FUNC SOURce STEP DC RAMPARC To set inquiry about the RAMP state for DCW t...

Страница 63: ...imit 50 0G 0 is OFF Data accuracy 1 0 M Data unit M Example Set the upper resistance for IR test in STEP1 to 1 M Command message FUNC SOUR STEP 1 IR UPPR 1 Query message FUNC SOUR STEP 1 IR UPPR Retur...

Страница 64: ...IR RTIM Data time Data format float Data range 0 999 0s 0 is OFF Data accuracy 0 1s Data unit s Example Set the rise time for IR test in STEP1 to 1s Command message FUNC SOUR STEP 1 IR RTIM 1 Query m...

Страница 65: ...SOUR STEP sn PA MESSAge Data message Data format characters Example Set the PA message in STEP1 to HELLO Command message FUNC SOUR STEP 1 PA MESSAge HELLO Query message FUNC SOUR STEP 1 PA MESSAge Ret...

Страница 66: ...Query message FUNC SOUR STEP sn OS SHOT Data rate Data format integer Data range 0 100 500 0 is OFF Data accuracy 10 Data unit Example Set the SHOT rate for OS test in STEP1 as 100 Command message FUN...

Страница 67: ...mode Query message FUNC BREAK STEPMODE Data step mode Data format integer Data range 0 1 0 normal step 1 voltage step Example Set the step mode for breakdown test to 1 voltage step Command message FU...

Страница 68: ...NC BREAK MODE Data mode Data format integer Data range 0 1 0 AC 1 DC Example Set the mode for breakdown test 1 DC Command message FUNC BREAK MODE 1 Query message FUNC BREAK MODE Return value 1 FUNC BR...

Страница 69: ...ry message FUNC BREAK UPPC Data current Data format float Data range 0 0001 25mA DC 0 001 100mA AC Data accuracy 0 0001mA DC 0 001mA AC Data unit mA Example Set the upper current for breakdown test to...

Страница 70: ...ue 1 0 FUNC BREAKdown RTIM To set inquiry about the RISE TIME for voltage breakdown test Syntax Command message FUNC BREAK RTIM time Query message FUNC BREAK RTIM Data time Data format float Data rang...

Страница 71: ...rc upper current for breakdown test to 5mA Command message FUNC BREAK ARC 5 Query message FUNC BREAK ARC Return value 5 0 FUNC BREAKdown CONTinue To set inquiry about the CONTINUE ON OFF for voltage b...

Страница 72: ...quiry about the TRIGGER DELAY for test Syntax Command message SYSTem MEA TRGDLY time Query message SYSTem MEA TRGDLY Data time Data format float Data range 0 0 99 9s Data accuracy 0 1s Data unit s Exa...

Страница 73: ...message SYSTem MEA RPTINT time Query message SYSTem MEA RPTINT Data time Data format float Data range 0 0 99 9s Data accuracy 0 1s Data unit s Example Set the test interval to 1 0s Command message SYS...

Страница 74: ...A STEPHOLD time Query message SYSTem MEA STEPHOLD Data time Data format float Data range 0 1 99 9s KEY Data accuracy 0 1s Data unit s Example Set the step hold interval to 1 0s Command message SYSTem...

Страница 75: ...cters Data range OFF 0 ON 1 Example Set the auto rang to ON Command message SYSTem MEA AUTORANGE 1 Query message SYSTem MEA AUTORANGE Return value 1 SYSTem MEA GFI To set inquiry about the state for G...

Страница 76: ...NV BEEPVOL 3 Query message SYSTem ENV BEEPVOL Return value 3 SYSTem ENV PASSVOL To set inquiry about the state for PASS VOLUM Syntax Command message SYSTem ENV PASSVOL ON OFF or 1 0 Query message SYST...

Страница 77: ...ate Query message SYSTem ENV KEYLOCK Data state Data format characters Data range 0 1 0 manual 1 bus Example Set key lock state to manual Command message SYSTem ENV KEYLOCK 0 Query message SYSTem ENV...

Страница 78: ...sage SYSTem ENV TIME hour minute second Query message SYSTem ENV TIME Data hour minute second Dataformat characters Data range 0 0 0 23 59 59 Example Set the system time to 16 23 23 Command message SY...

Страница 79: ...al storage Syntax Command message MMEM COPY file name Data file name Data format characters Example Set copy file name to TH9110TEST Command message MMEM COPY TH9110TEST Return value copy pass return...

Страница 80: ...copy fail return ERROR 5 2 7 FETCH Subsystem Commands The FETCh subsystem command is used to obtain the measurement results of the tester and is issued after the test is started until the end of the...

Страница 81: ...Note The tester will automatically return each test result by default per step test result 5 2 8 Other Commands IDN Inquiry about the model and version information of the instrument Return message ma...

Страница 82: ...0mA voltage 1 5kV Resolution 0 1uA Accuracy 1 reading 0 5 full scale Maximum Short Circuit Current 200mA AC test only Fast Discharge Function Auto discharge after test ends DCW Insulation Resistance T...

Страница 83: ...hen both ends are short circuited high voltage delivery is allowed refer to Figure3 2 INTER LOCK Panel Operation Protection Key Lock Password Alarm Indication Pass short beep green light Fail long bee...

Страница 84: ...onal and technical personnel to carry out please do not replace the internal components of the tester without authorization after the tester is repaired it needs to be re calibrated to avoid affecting...

Отзывы: