ー 30 ー
5-9. Operation completion
The buzzer sounds three times when the set sterilization time has elapsed and the sterilizing
process is completed.
When the pressure in the chamber falls to atmospheric pressure, the color of the operation
indication lamp changes from red to yellow.
When the temperature inside the chamber or temperature measured by the temperature
sensor for materials to be sterilized (optional) falls below a preset value to release the lid
interlock, the buzzer will sound six times and the lid will unlock and open.
When the temperature inside the chamber or temperature measured by the temperature
sensor for materials to be sterilized (optional) falls to 60°C and the process of temperature
decreasing ends, the buzzer will sound ten times to indicate an operation is completed.
After completing the operation, press the [Stop] key. The lid indication lamp lights up and
the lid open/close detection system is released.
※
Operation can be interrupted by pressing the [Stop] key while running.
※
When operation is interrupted, the status shifts to the temperature and pressure reduction process
or temperature reduction process.
※
The lid open/close detection system is activated by opening or closing the door.
①
②
③
④
⑤
The completion indication lamp blinks when the temperature in the chamber falls to that which
releases the interlock of the chamber lid allowing the chamber lid to open. It is recommended
that the chamber lid be opened after the temperature in the chamber falls to 60 °C and the
operation indication lamp goes out because there is likelihood of scalds due to sudden boiling
immediately after operation.