ー 37 ー
Hold down the [Function] key when the
operation indication lamp is not illuminated.
After approximately two seconds, “F01”
blinks at the temperature display.
Display “F02” by pressing the [Temperature
setting] keys.
The temperature releasing the interlock of the
chamber lid is displayed at the time display.
Press the [Start] key.
The time display blinks.
Press the [Time setting] keys to display the desired temperature releasing the interlock of the
chamber lid.
Press the [Start] key to determine the temperature which releases the interlock of the chamber
lid.
※
The temperature is set to 97 °C at the time of shipment from the factory.
※
The settable range of the temperature which releases the interlock of the chamber lid is 60 to 100 °C.
<
Memorizing the preferred course
>
①
②
③
④
⑤
The autoclave lid interlock system does not allow the lid to be opened while the temperature
inside the chamber or temperature measured by the temperature sensor for materials to be
sterilized (optional) is high. The temperature to release the lid interlock can be changed
according to the operating conditions.
Hold down the [Function] key when the operation indication lamp is not illuminated.
After approximately two seconds, “F01” blinks at the temperature display.
The warming time is displayed at the time display.
Press the [Start] key.
The time display blinks.
Press the [Time setting] keys to display the desired warming time.
Press the [Start] key to determine the warming time.
※
The temperature is set to 4 hours at the time of shipment from the factory.
※
The settable range of the warming time is 1 to 99 hours.
①
②
③
④
6-8. Setting the warming time
The warming time can be changed.
6-7.
Setting the temperature which releases the interlock of the chamber lid