ー 48 ー
WARNING
CAUTION
After replacing the chamber lid gasket, be sure to make a test run.
●
Make sure the chamber lid gasket is properly installed, otherwise a steam leak
through the lid may cause burns.
Extra care should be taken when handling the spring band of the
gasket gland.
●
The elastic force may cause unexpected injuries. Be sure no other people are in
the immediate vicinity when handling the spring band.
If the chamber lid gasket shows any signs of deterioration or damage, replace it as follows: Use
extreme caution while performing the replacing procedure and ensure there is no damage to the
chamber lid gasket or the spring band of the gasket gland.
Use only replacement lid gasket parts compatible with the model in use.
SX-300E, SX-500E
Lid gasket parts kit 500
SX-700 E
Lid gasket parts kit 700
8-3. Replacing the chamber lid gasket
1. Remove the chamber lid gasket.
Grip the upper right of the lid gasket
with the hands as shown in the photo
and pull it down to the lower left to
make a deflection on both the lid gasket
and the spring band of the gasket gland.
Pull both the lid gasket and the spring
band of the gasket gland off the rim to
remove them.
<