ー 26 ー
Place articles to be sterilized in the stainless basket
normally attached or an optional stainless bucket.
※
Be careful not to hit articles to be sterilized on the
lid gasket or the temperature sensor when placing or
removing the articles. They may be damaged.
Place articles gently in the chamber.
※
Place articles so that they do not come above the
upper level of the lid hooks. When the chamber lid
is closed with articles placed above this level, they
may be damaged.
Close the chamber lid.
<
Placing articles to be sterilized
>
①
②
③
5-6. Placing articles to be sterilized
WARNING
Do not operate the autoclave with flammable or explosive articles in
place.
●
Contravention is dangerous with likelihood of fire or explosion.
Do not sterile sealed articles.
●
Contravention is dangerous with likelihood of burns
or serious accident due to blowout during removal
of articles. In addition, contravention is dangerous
with likelihood of damage to parts, or burns or serious
accident due to explosive breakage because the pressure
in the chamber may rise abnormally.
Do not sterilize glass equipment having cracks or
scratches.
●
Contravention is dangerous with likelihood of burns
or serious accident due to blowout during removal of
equipment.
Do not block the hole in the inside of the chamber with articles to be
sterilized.
●
Contravention is dangerous with likelihood of serious accident such as blowout
because the pressure in the chamber cannot be controlled when the hole at the
inside of the chamber is blocked with an article such as a sterilizing bag. Make
sure that articles are placed in the basket provided etc. so that the vicinity of the
hole is not blocked.
Upper level of lid hook
lid hook