5.4 Routine maintenance
This section details the daily care of MR-6000 itself. Clean each devicie connected by this
instrument in accordance with each operation manual.
■
Be sure to turn off the relevant devices and disconnect the power
plug from the outlet to ensure safety before starting cleaning.
■
Do not use organic solvents such as thinner, benzene, or acetone.
These solvents may damage the surface of the instrument.
■
Do not wipe or strongly rub the measurement window when dust
etc. is adhered. Optical components may be damaged.
5.4.1 Measurement window
The measurement window protrudes from the surface of the instrument and move
forward an eye when taking measurements. Therefore, the measurement window is
easily contaminated with dirt, dust, tears, etc. Clean the measurement window as
needed according to the following procedure.
1) Use a commercial blower brush etc. on the measurement window to remove dust.
2) Without rubbing, lightly wipe the glass section with a swab or soft cloth dampened
with a small amount of ethanol
for disinfection
.
5.4.2 Forehead pad and chinrest
Without rubbing, gently wipe sections that touch patients directly with a cloth
dampened with ethanol
for disinfection
before starting measurement for a new patient.
5.4.3 External surface and touch panel of the instrument
Use soft, dry cloth to clean the instrument. When it is very dirty, clean it using a well
wrung cloth dampened with diluted neutral detergent, and then wipe it with a dry cloth.
When the surface of the touch panel is dirty, wipe it with a soft cloth.
Содержание MR-6000
Страница 2: ......
Страница 24: ... 1 18 This page is intentionally left blank ...
Страница 30: ... 2 6 This page is intentionally left blank ...
Страница 46: ... 4 2 This page is intentionally left blank ...
Страница 56: ... 6 4 This page is intentionally left blank ...
Страница 58: ... 7 2 This page is intentionally left blank ...
Страница 67: ...8 9 This page is intentionally left blank ...
Страница 68: ......
Страница 70: ...0120 ...