3-12
measurement results into consideration when using this
instrument.
When using the intraocular pressure adjustment function, set
appropriate parameters for the adjustment formula. The adjusted
intraocular pressure to be calculated is a reference value.
Explain to the patient that a pull of air will be expelled when
measuring intraocular pressure to avoid surprise upon
measurement.
<Measurement procedures>
1) Switch the measurement mode to "Tono/Pachy."
* Touch the “Tono” button when
measuring intraocular
pressure only and the “Pachy” button when measuring
corneal thickness only.
2) Perform alignment and take measurements with the joystick
or the joystick button on the measurement screen. If Auto
Shot is turned on, measurement starts automatically after
alignment is complete. The corneal thickness is measured
first, followed by measurement of intraocular pressure.
3) Once data of the intraocular pressure and corneal thickness
is obtained for the specified number of measurements,
measurement completes.