1-5
■
■
Turn power off and place the dust cover over the instrument when not
being used. The measurement accuracy will deteriorate significantly if
the optical section in the instrument becomes dirty. Disconnect the
power plug from the outlet when not being used for long periods. Dust
may accumulate on the plug, resulting in fire.
1.1.2 Installation
■
Do not install this instrument near a location where flammable
substances, explosives or chemicals are used or stored. Any chemicals
or vapor entering the instrument may result in fire or explosion.
■
Do not use the instrument near equipment with a strong
electromagnetic wave disturbance source (MRI, high frequency
electrosurgical device, etc.).
A strong magnetic field may cause noise
and affect measurement.
■
Install the instrument in a location free of water or chemicals. Any water
or chemicals entering the instrument may cause an electric shock or
failure.
■
Install the instrument in a location where the power cord can be
connected and disconnected easily.
Содержание MR-6000
Страница 2: ......
Страница 24: ... 1 18 This page is intentionally left blank ...
Страница 30: ... 2 6 This page is intentionally left blank ...
Страница 46: ... 4 2 This page is intentionally left blank ...
Страница 56: ... 6 4 This page is intentionally left blank ...
Страница 58: ... 7 2 This page is intentionally left blank ...
Страница 67: ...8 9 This page is intentionally left blank ...
Страница 68: ......
Страница 70: ...0120 ...