DIGITAL WEIGHING INDICATOR TI-5700
- 15 -
5.3. Multi calibration (5 weighing point)
Even if the single calibration has been completed, there may still remain a linearity
deviation caused by the performance of the weighing unit.
The multi calibration can rectify or reduce the linearity deviation using weighing
points during the zero and capacity setting.
Up to five weighing points can be specified.
Step
Operation & Description
Display
1
How to enter the calibration mode:
Press the CAL key which is located behind the CAL cover at
the front panel of indicator.
When
“
CAL
”
is displays, press the NET/GROSS key.
Then this calibration mode is stared.
2
Setting maximum capacity:
After
“
CAPA
” is displayed, input the max. capacity with keys.
(
>
: increase the digit,
∧
: increase the value)
Press ENTER key to save and move to next step.
3
Setting minimum scale division:
After
“
DIV
” is displayed, select min. scale division with keys.
(
>
: increase the value,
∧
: decrease the value)
Press ENTER key to save and move to next step.
If you press the ESC key, it turns back to weighing mode after
save maximum capacity & minimum scale division.
The display resolution is cannot be over than 50,000.
(Display resolution = Max. capacity / Min. scale division)
4
Zero calibration:
Place nothing on the weighing unit.
When
“
ZERO
” is displayed, press the ENTER key after the
Stable mark has turned on.
If you press the ESC key, it turns back to weighing mode after
execute zero calibration.
Do not press the keys while the stable mark is off.
5
Input 1
st
balance weight:
After
“
RANGE1
” is displayed, input 1
st
balance weight with
keys. (
>
: increase the digit,
∧
: increase the value)
Press ENTER key to save and move to next step.