DIGITAL WEIGHING INDICATOR TI-5700
- 48 -
◊
Format 2 (22 bytes
– CAS company standard)
Header 1
Header 2
ID Data
Unit Terminator
S T , G S ,
,
0 . 0
k g C
R
L
F
Lamp status
Header 1
ST
Stable weight data (0x53) (0x54)
US
Unstable weight data (0x55) (0x53)
OL
Overload (0x4F) (0x4C)
HD
Hold weight data (0x48) (0x44)
Header 2
GS
Gross data (0x47) (0x53)
NT
Net data (0x4E) (0x54)
ID
Command address ID is selected in F35, F44.
(If the ID is 13, 0x13 is sent)
Data
Example 1) 13.5kg
„ ‟, „ ‟, „ ‟, „ ‟, „1‟, „3‟, „.‟, „5‟
Example 2) 135kg
„ ‟, „ ‟, „ ‟, „ ‟, „1‟, „3‟, „5‟, „ ‟
Example 3) -13.5kg
„-‟, „ ‟, „ ‟, „ ‟, „1‟, „3‟, „5‟, „ ‟
Lamp
statue
Bit7
Bit6
Bit5
Bit4
Bit3
Bit2
Bit1
Bit0
1
Stable
1
Hold
1
Net
Tare
Zero
Unit
kg
(0x6B) (0x67)
g
(0x20) (0x67)
lb
(0x6C) (0x62)
Terminator
C
R
L
F
(0x0D) (0x0A)