Introduction 

The

 

ProComp Infiniti™ encoder is an eight (8) channel, multi-modality device for real-time 

computerized biofeedback and data acquisition. It has 8 protected pin sensor inputs with two 
channels sampled at 2048 s/s and six channels sampled at 256 s/s. The ProComp Infiniti encoder 

is able to render a wide and comprehensive range of objective physiological signs used in clinical 

observation and biofeedback. 

 

 

 

The ProComp Infiniti basic package includes the following hardware: 

ƒ

 

One eight-channel ProComp Infiniti encoder unit 

ƒ

 

One TT-USB interface unit 

ƒ

 

A supply of fiber optic cable (1’ and 10’ cables) 

ƒ

 

Four alkaline ‘AA’ batteries 

 

Thought Technology’s advanced design and active electronic sensors meet exacting standards for 

instrument accuracy, sensitivity, durability, and ease of use.  All sensors are completely non-

invasive and require little or no preparation for use.  Depending on the software being used with 
the ProComp Infiniti encoder, these may include sensors specialized for: 

ƒ

 

Electromyography (EMG) 

ƒ

 

Electrocardiography (EKG) 

ƒ

 

Electroencephalography (EEG) 

ƒ

 

Skin Conductance 

ƒ

 

Skin Temperature  

ƒ

 

Voltage Isolation 

ƒ

 

Respiration waveform, rate, and amplitude 

ƒ

 

Blood Volume Pulse waveform, heart rate, and 
amplitude (BVP) 

ƒ

 

Goniometer 

ƒ

 

Force 

ƒ

 

Torsiometer 

 

The sensors pass signals to the host computer via the microprocessor-controlled ProComp Infiniti 

encoder unit.  The encoder samples the incoming signals, digitizes, encodes, and transmits the 

sampled data to the TT-USB interface unit.  A fiber optic cable is used for transmission, providing 

maximum freedom of movement, signal fidelity, and electrical isolation. Cable lengths up to 25 

feet can be used without signal degradation.  A unique feature of the system design allows inputs 
to accept any sensor, interchangeably (except the raw EMG MyoScan sensors which are used on 

channel A and B). This allows a wide variety of configurations to be created by simply changing 

the sensor types.   

 

The TT-USB interface unit is connected to one of the host computer’s USB ports. It receives the 

data arriving from the encoder in optical form and converts it into the USB format to 

communicate with the software. 

 

Please note, some hardware features may not be supported by all software programs.  Consult the 
software manual for a full list of features supported. 

ProComp Infiniti Hardware Manual 

8

Содержание Procomp Infiniti

Страница 1: ...niti Hardware Manual Thought Technology Ltd 2180 Belgrave Avenue Montreal QC H4A 2L8 Canada Tel 800 361 3651 514 489 8251 Fax 514 489 8255 E mail mail thoughttechnology com Webpage http www thoughttec...

Страница 2: ...ProComp Infiniti Hardware Manual 2...

Страница 3: ...8 Product Name ProComp Infiniti System Product T7110M T7120M T7150M T7160M T7170M T7500M T7520M Device Name ProComp Infiniti Encoder Device SA7500 EC REP EMERGO EUROPE Molenstraat 15 2513 BH The Hague...

Страница 4: ...Pro sensor Be sure to keep in mind that a very relaxed muscle should provide an EMG reading of approximately 1 3 V Explosion Hazard Do not use in the presence of a flammable anesthetic mixture with a...

Страница 5: ...e and or frequency domain while the device is being used for biofeedback or other purposes If anomalies are observed on acquired signals and if you suspect a problem with electromagnetic interference...

Страница 6: ...2 SURFACE ELECTROMYOGRAPHY SEMG 22 Sensor Types 22 Operating Principle 22 Measurement Unit 23 Sensor Placement 23 Typical Signal 23 Zeroing 24 ELECTROCARDIOGRAPHY EKG 24 Sensor Type 24 Operating Princ...

Страница 7: ...ensor Type 34 Operating Principle 34 Measurement Unit 34 FORCE ADAPTER 35 Sensor Type 35 Operating Principle 35 Measurement Unit 35 SELF CALIBRATION 36 CONDITION CODE SUMMARY 37 REFERENCE 38 TECHNICAL...

Страница 8: ...EEG Skin Conductance Skin Temperature Voltage Isolation Respiration waveform rate and amplitude Blood Volume Pulse waveform heart rate and amplitude BVP Goniometer Force Torsiometer The sensors pass s...

Страница 9: ...Before inserting the cable it may be necessary to loosen the jack heads by turning them counter clockwise 3 Ensure that both ends of the optical cable are inserted all the way about of an inch 1 9 cm...

Страница 10: ...In Windows XP screens similar to the one below may appear Windows should recognize when new device is plugged into one of its USB ports Let windows install the software automatically If the BioGraph I...

Страница 11: ...ve a new feature that controls the installation of unsigned device drivers There are three settings for Driver Signing Block This will prevent the installation of unsigned drivers Warn This will promp...

Страница 12: ...Windows 2000 Make sure that either Warn or Ignore is selected Then click on OK ProComp Infiniti Hardware Manual 12...

Страница 13: ...Channels A B 2048 samples second Channels C to J 256 samples second 1 4 3 2 ON 0 1 For use with the BioGraph CardioPro or MultiTrace software and the Pro SB interface Channels A B 256 samples second...

Страница 14: ...e ProComp Infiniti it is better to connect only the sensors that are going to be used for a session before you start recording this will ensure maximal battery life Most Thought Technology software ap...

Страница 15: ...stem other than FAT will render it Inoperable Users should follow the appropriate procedures used for disconnecting peripheral devices when unplugging the Compact Card from its PC reader Compact Flash...

Страница 16: ...ns Then select Set Compact Flash Date Time Specify a date and time a brief period in the future when you will insert the Compact Flash in the encoder Press OK and wait for the operation to complete If...

Страница 17: ...mand If no space remains on the Compact Flash only the initial 1 second blink is performed This sequence is replayed when the dipswitch is changed Memory Remaining Initial 1 Sec 0 10 20 30 40 50 60 70...

Страница 18: ...or 3 sec Recording Stops White LED flashes to show remaining space when encoder is switched on S 1 CF Recording is the white LED turns on First the blue LED will tur encoder to turn off After the whit...

Страница 19: ...natural electrical field In most cases you can do so by fastening the belt clip to the client s waist Caring For The Fiber Optic Cable A fiber optic FO cable is used for transmitting the sensor data...

Страница 20: ...ip On the TT USB Interface you will also notice a hole labeled Switch This plug is for connecting external devices such as a muscle stimulation device when you want to use the switch control feedback...

Страница 21: ...ensor can be placed on the client you must snap on a set of disposable electrodes Three types of electrodes are available from Thought Technology Triodes Single Strip and Uni Gels The Triode and Singl...

Страница 22: ...nector itself or the extender cable s rounded snap SURFACE ELECTROMYOGRAPHY SEMG Sensor Types The MyoScan and the MyoScan Pro are the surface EMG SEMG sensors for the ProComp Infiniti device SEMG is r...

Страница 23: ...produce microvolt readings of up to 1600 V To properly monitor their activity set the sensor to the 1600W position The MyoScan sensor s active range is from 10 to 500 Hz It can record SEMG signals fr...

Страница 24: ...t always connect the same sensor to the same encoder input when you use that screen Remember to remove the zeroing plug after adjusting the MyoScan Pro sensor s offset NOTE Zeroing is not required for...

Страница 25: ...oo invasive it is possible to place the electrodes on the client s forearms For a forearm placement the negative lead should be placed on the right forearm while both the positive and ground leads are...

Страница 26: ...he EEG Flex Pro sensor measures microvolts V raw A normal EEG signal recorded from the scalp will have an amplitude between 0 1 and about 200 V The raw EEG signal is mainly used to evaluate the qualit...

Страница 27: ...placed on the earlobe of the opposite side Note bipolar is not the same as bilateral A bipolar placement provides only one signal A bilateral placement requires two sensors and gives two signals When...

Страница 28: ...e have the same head size and shape the first step in defining correct electrode locations is to make some measurements Two measures are required The distance between the supra nasal grove nasion and...

Страница 29: ...C3 Fz C4 and Pz Measuring a straight line between these points is acceptable Finally calculate the distances between Fpz and T3 Fpz and T4 Oz and T3 and Oz and T4 Calculate 10 on either sides of Fpz a...

Страница 30: ...easures S and m are equivalent Normal readings for skin conductance in a relaxed state are around 2 S but readings can vary greatly with environmental factors and skin type Sensor Placement The skin c...

Страница 31: ...changes are fairly slow and do not have much amplitude To monitor temperature changes effectively it is necessary to have a fairly small vertical scale size BLOOD VOLUME PULSE Sensor Type To record bl...

Страница 32: ...ill increase and decrease with changes in sympathetic arousal RESPIRATION AMPLITUDE RESP FLEX PRO Sensor Type The Respiration sensor is labeled Respiration Flex Pro Although the respiration sensor is...

Страница 33: ...be allergic to rubber Please verify your client s allergic conditions before using the respiration sensor When using two sensors strap one around the client s chest just above the breasts and one at...

Страница 34: ...Unit A goniometer sensor converts its position to a raw value as the position changes the value corresponding to that position changes with it Most goniometers have a specified minimum and maximum def...

Страница 35: ...d changes with it This raw value can be transformed into a meaningful calibrated value in the software by using the linear transform virtual channel This linear transform takes the minimum and maximum...

Страница 36: ...op flashing and light up steadily as before If the blue LED flashes quickly for about 1 second before lighting up steadily this indicates that the self calibration failed If this happens repeatedly wh...

Страница 37: ...second then steady on Failed self calibration WHITE Long blink about 1 second then flashes 1 to 10 times Compact flash recognized and ready to record with each flash indicating 10 of remaining capacit...

Страница 38: ...ess Provide a detailed description of the problem you are experiencing and your telephone fax number and email see form on the last page of this manual In the USA ship insured to Thought Technology Lt...

Страница 39: ...CABLE FIBRE OPTIC 1FT SA9451 CABLE FIBRE OPTIC 10FT SA9480 CABLE FIBRE OPTIC 15FT SA9452 CABLE FIBRE OPTIC 25FT SA3403 ELECTRODE EXTENDER CABLES THREE PACK SA8950 PRO SB SA8952 PRO SB DUAL T8720M SENS...

Страница 40: ...Placing Orders Outside USA Tel 514 489 8251 Fax 514 489 8255 In USA Toll Free Tel 1 800 361 3651 E Mail mail thoughttechnology com 40...

Страница 41: ...2000 VRMS Sensitivity 0 1 VRMS CMRR 130dB Channel Bandwidth 10Hz 1kHz Signal Output Range 0 1 0V RMS Input Output Gain 500 Supply Voltage 7 26V 0 02V Current Consumption 0 7mA 0 25mA Accuracy 0 3 VRMS...

Страница 42: ...mpedance 1 000 000M in parallel with 10pF Input Range 0 200 VRMS Sensitivity 0 1 VRMS Bandwidth 2Hz 1kHz CMRR 130dB Accuracy 0 3 VRMS 5 of reading 10 C to 40 C Temperature Sensor SA9310M Length Approx...

Страница 43: ...0 30 0 S Accuracy 5 and 0 2 S Goniometer Adapter Hardware Specifications Size approx 370mm x 370mm x 100mm 1 45 x 1 45 x 0 44 Weight approx 15g 0 5 oz Input Impedance 1M Signal Input Range 180 180 5...

Страница 44: ...nd DC 64Hz 200 samples second DC 8Hz 32 samples second DC 8Hz 20 samples second Anti aliasing Filter 5th order Butterworth Alias rejection 30dB typical DC gain accuracy 0 5 initial or after self calib...

Страница 45: ...Fiber Optic Cable 9 Force Adapter 35 G Goniometer Adapter 34 I Interface Unit Connecting 9 M MyoScan 22 MyoScan Pro 22 O Order Placing 38 P PROCOMP Basic package 8 ProComp Encoder Setup 13 Protected...

Страница 46: ...received We will repair or replace your unit s free of charge This warranty does not apply to damage incurred through accident alteration or abuse nor to sensor damage created by static electricity Do...

Страница 47: ...ing any equipment Remove or copy this form and include it with the unit s Include copy of original invoice and return to the address in the Returning Equipment section Name Company Address Phone No Fa...

Страница 48: ...property contained herein is confidential between Thought Technology Ltd and the client and remains the exclusive property of Thought Technology Ltd If you find any problems in the documentation plea...

Отзывы: