BP209 Series
7 Operation Instructions
Rev: 8.1, 14-Feb-2022
© 2022 Thorlabs
Page 73
It is recommended that the bandwidth be set manually. The guideline is the same as for CW
signals - the smaller the beam diameter, the higher the bandwidth is required: Small beam dia-
meters cause short photo diode current pulses when the slit is scanning it. If the BW is set too
low, the rising edge of the photo diode current is delayed, and the falling edge overshoots. This
leads to an artificially lower peak amplitude.
The X profile of a 130 µm beam illustrates this:
Bandwidth - 125 kHz
Bandwidth - 16 kHz
At larger beam diameters, higher bandwidth results in a higher noise level - as can be seen for
a beam with of about 3 mm:
Bandwidth 16 kHz
Bandwidth - 125 kHz
Gain Index
For pulsed lasers, it is recommended that the gain index be set manually in order to avoid sat-
uration of the amplifier.
Содержание M2MS
Страница 2: ...Copyright 2022 Thorlabs Version Date 8 1 14 Feb 2022 ...
Страница 73: ...BP209 Series 7 Operation Instructions Rev 8 1 14 Feb 2022 2022 Thorlabs Page 71 Example ...
Страница 92: ...BP209 Series 7 Operation Instructions Page 90 21424 D02 ...
Страница 119: ...BP209 Series 7 Operation Instructions Rev 8 1 14 Feb 2022 2022 Thorlabs Page 117 ...
Страница 151: ...BP209 Series 11 Appendix Rev 8 1 14 Feb 2022 2022 Thorlabs Page 149 Typical Responsivity Extended InGaAs Photodiodes ...
Страница 154: ...BP209 Series 11 Appendix Page 152 21424 D02 11 7 Drawing BP209 Drawing BP209 Series Imperial Version ...
Страница 155: ...BP209 Series 11 Appendix Rev 8 1 14 Feb 2022 2022 Thorlabs Page 153 Drawing BP209 Series Metric Version ...
Страница 157: ...BP209 Series 11 Appendix Rev 8 1 14 Feb 2022 2022 Thorlabs Page 155 11 9 Drawing BP209 Mounting Adapter ...
Страница 158: ...BP209 Series 11 Appendix Page 156 21424 D02 11 10 Drawing M2MS BP209 ...
Страница 162: ...BP209 Series 11 Appendix Page 160 21424 D02 11 14 Certifications and Compliances ...
Страница 167: ...www thorlabs com ...