BP209 Series
7 Operation Instructions
Rev: 8.1, 14-Feb-2022
© 2022 Thorlabs
Page 75
Preparation
·
Set Gain Mode to "
Automatic
".
·
Set X and Y Slit Bandwidth to "
1000 kHz
".
·
Set Base Line Mode to "
Manual
".
·
Set Photodiode Bias to "
On
" - this shortens the
rise/fall time of the photo diode.
·
Set Aperture Width to "
0.9 mm
" - the software
uses only an area of Ø 0.9 mm in the sensor center
for calculation.
·
Set Active Slit Pair to "
25 µm
".
·
Set Scanning Mode (for alignment) to "
Slit
".
·
Set Scan Rate to "
2 Hz
".
·
Enable Scan Rate Correction.
Enable X and Y profiles.
Disable Gaussian and Bessel fits and zoom in the profiles. A proper alignment is reflected in a
maximum amplitude and a flat top of the profile. The flat top is a result of the small beam size
Содержание M2MS
Страница 2: ...Copyright 2022 Thorlabs Version Date 8 1 14 Feb 2022 ...
Страница 73: ...BP209 Series 7 Operation Instructions Rev 8 1 14 Feb 2022 2022 Thorlabs Page 71 Example ...
Страница 92: ...BP209 Series 7 Operation Instructions Page 90 21424 D02 ...
Страница 119: ...BP209 Series 7 Operation Instructions Rev 8 1 14 Feb 2022 2022 Thorlabs Page 117 ...
Страница 151: ...BP209 Series 11 Appendix Rev 8 1 14 Feb 2022 2022 Thorlabs Page 149 Typical Responsivity Extended InGaAs Photodiodes ...
Страница 154: ...BP209 Series 11 Appendix Page 152 21424 D02 11 7 Drawing BP209 Drawing BP209 Series Imperial Version ...
Страница 155: ...BP209 Series 11 Appendix Rev 8 1 14 Feb 2022 2022 Thorlabs Page 153 Drawing BP209 Series Metric Version ...
Страница 157: ...BP209 Series 11 Appendix Rev 8 1 14 Feb 2022 2022 Thorlabs Page 155 11 9 Drawing BP209 Mounting Adapter ...
Страница 158: ...BP209 Series 11 Appendix Page 156 21424 D02 11 10 Drawing M2MS BP209 ...
Страница 162: ...BP209 Series 11 Appendix Page 160 21424 D02 11 14 Certifications and Compliances ...
Страница 167: ...www thorlabs com ...