5.5 Troubleshooting
LDC2xx / page 28
5.5 Troubleshooting
In case that your LDC200 system show s malfunction please check the following
items:
Module does not work at all (no display on the mainframe):
Mainframe LDC200 connected properly to the mains?
Connect the LDC200 to the power line paying attention to the right voltage
setting of your mainframe.
Mainframe LDC200 turned on?
Turn on your LDC200 with the key mains-switch.
Control the fuse at the rear panel of the LDC200 mainframe.
If blown up replace the fuse by the correct type
You don’t get the desired laser output power
Is the interlock closed?
Control the resistance between the interlock pins of the connector jack not
to be more than 430
Ω
.
(refer to section 3.1.2, “Interlock and control LED for LASER ON” on page
11)
Do you have turned on the laser output with the "ON" button?
Change the setting from “off” to “on”.
The LED “ON” on the front panel of the mainframe must be on
Is the current limit I
LIM
set to 0?
Adjust the hardware limit I
LIM
by means of the potentiometer on the LDC2xx
front panel to an appropriate value.