3.4 Constant current mode (CONST I)
LDC2xx / page 17
3.4 Constant current mode (CONST I)
Switch on the LDC2xx.
Select a suitable current limit "I
LIM
" (refer to chapter 3.2).
Select the appropriate laser polarity (refer to chapter 3.3).
Connect the laser diode (refer to chapter 3.1.3).
Select the display "I
LD
" with the button (L18) or (L19).
Set the adjust knob (L10) completely to the left.
Select constant current mode with key "I" (L23).
Switch on the output by pressing key "ON" (L9). The LED "ON" (L8) lights up, the
output is activated and the curr ent slowly rises (about 1 s) to the set value with knob
(L10).
NOTE:
The current output can only be switched on if the jack "LD OUTPUT" (R5)
is connected correctly.
The display shows the injection current "I
LD
".
With the adjust knob (L10) the laser curr ent can be set continuously between 0 mA
and the selected current limit "I
LIM
".
If the injection current "I
LD
" reaches the set current limit "I
LIM
" during operation the
LED "LIM" (L7) lights up, a short beep as
warning signal is heard and the laser
current is limited to the value of the current limit "I
LIM
".
If the LED "LIM" (L7) lights up noise
and ripple do not longer correspond to the
specifications for normal operation (refer to chapter 6.3). However exceeding the set
current limit "I
LIM
" is not possible.