Instrument specifications
Instrument dimensions and specifications
Table 7
Specification
iBright
™
CL1000 Instrument iBright
™
FL1000 Instrument
Dimensions (L × W × H)
68.0 × 38.2 × 59.8 cm
Weight
52 kg (115 lbs)
Instrument clearance (Back)
15.25 cm (6 in.)
Electrical requirements
WARNING!
For safety, the power outlet used for powering the instrument
must be accessible at all times. In case of emergency, you must be able to
immediately disconnect the main power supply to the instrument. Allow
adequate space between the wall and the equipment so the power cord can be
disconnected in case of emergency.
• Electric receptacle with grounding capability
• Maximum power dissipation: ~960 W (not including computer and monitor)
• Mains AC line voltage tolerances must be up to ±10 percent of nominal voltage
Device
Rated voltage
Circuit
required
Rated
frequency
Rated
power
iBright
™
CL1000 and
FL1000 Instruments
100–240 ±10% VAC
[1]
10 A
50/60 Hz
410 W
[1]
If the supplied power fluctuates beyond the rated voltage, a power line regulator may be required. High or low
voltages can adversely affect the electronic components of the instrument.
A
iBright
™
Imaging Systems User Guide
31