Nucleic Acid Gels
Mode to image DNA and RNA gels stained with fluorescent nucleic acid stains such
as Ethidium Bromide, SYBR
™
Safe, SYBR
™
Green (I or II), SYBR
™
Gold and similar
products.
1.
Touch Smart exposure. Alternatively, to set a manual exposure time, touch the
center of the dial.
The imager acquires a series of short exposures and then renders a preview
image and a recommended exposure time. This is not a real image.
2.
To adjust or set a manual exposure time, select a method:
• Touch or within dial.
• Finger swipe in the segmented dial.
• Touch the dial center box to select a preset exposure time or touch Custom to
manually input a time.
3.
Touch Enter.
The image preview will update in real-time.
4.
Touch Capture to acquire image with the indicated exposure time.
Captured image appears on screen and automatically saves to the gallery.
5.
If an acceptable image, touch Export, Gallery or Analyze. If an unacceptable
image, touch More options
4
Image adjust to optimize image or touch Trash to
remove image, then return to step 2 to adjust exposure conditions.
Image using
Nucleic Acid Gel
Operate the iBright
™
imager