Unlink account
Disconnects the selected user account from the cloud.
Reset PIN
Existing PIN for selected user is deleted and reset to 0000.
Run instrument diagnostics
Instrument functions can be checked by a user to confirm proper hardware operating
conditions.
1.
Touch .
2.
Touch Service Tools.
3.
Touch Instrument diagnostic.
4.
Touch Start test.
The instrument will check the instrument hardware for correct operation and
settings. The test will take ~10 minutes to complete.
Note:
Ensure the drawer is not blocked. The drawer will open as part of the
testing.
Clean the instrument
Repeated instrument use can cause spots and smudges on the sample stage, which
can then decrease image quality. Clean the sample stage as needed.
Materials required
• Safety glasses
• Powder-free gloves
• Tissue, lint-free
• Deionized water
• Ethanol, 70% solution
Note:
Avoid the use of detergents.
1.
Open the drawer to expose the sample stage.
2.
Lightly spray the glass surface with deionized water or a 70% ethanol solution.
3.
Wipe the surface with a lint-free tissue until sufficiently clean.
4.
Close the drawer and operate the instrument as normal.
Maintain the instrument
Run instrument diagnostics
28
iBright
™
Imaging Systems User Guide