Chemi Blots
Mode to image chemiluminescent Western blots.
After you choose the imaging mode, capture an image of the sample.
If you are using the Fluorescent Blots mode (FL1000 model only), select one or more
dyes (see “Select a dye in the Fluorescent Blots mode“ on page 13).
IMPORTANT!
To image a Coomassie or silver-stained gel using the Protein Gel mode,
place a white background under the sample.
1.
Touch Smart Exposure.
The imager acquires a series of short exposures, then renders a preview image
and recommends an exposure time. This is not a real image.
2.
(Optional) Adjust the exposure time using the following methods:
• Touch or .
• Finger swipe in the segmented dial.
• Touch the dial center box to select a preset exposure time or touch Custom to
manually input a time.
3.
Touch Enter.
The image preview updates in real-time.
4.
Touch Capture to acquire an image.
The captured image is displayed and automatically saves to the gallery.
5.
(Optional) Touch More options
4
Image adjust to optimize image or touch Trash
to delete the image.
Capture an image
of the sample
Operate the iBright
™
imager
Chemi Blots
12
iBright
™
Imaging Systems User Guide
Содержание Invitrogen iBright CL1000