Options available in modes
Each mode offers More Options on certain workflow screens to provide detailed
camera and image adjustment.
Table 3
Camera
Suboption
Detail
Resolution/Sensitivity
Changes binning setting to increase resolution/decrease
sensitivity or decrease resolution/increase sensitivity.
Zoom/Focus
Changes zoom level to increase or decrease image area.
Optimizes focus for sharpness.
Sample rotation
Imager mechanically rotates sample ±10° depending on
orientation. Rotation is automatic, but also allows ability to
adjust in More Options
4
Camera Settings.
Table 4
Routines
Suboption
Detail
Multi-exposure
Series of 5 preset exposure times producing one image for
each time period.
Signal accumulation
Series of exposures where user defines the first capture
(time interval), the last capture (cumulative capture time)
and the total number of captures.
Signal accumulation
Note:
Images acquired using Signal Accumulation are not
recommended for publication or analysis. This feature is
recommended for determining an optimal exposure time.
Table 5
Image adjust
Suboption
Detail
Contrast
Grayscale differentiation between image features. Can use
Auto contrast, Gamma, White slider bar, or Black slider
bar.
Channels and Layers
Allow user to navigate between individual channels
associated with a common image file; pertains only to
Chemi and Fluorescent blot modes (e.g., Channels:
membrane, Alexa Fluor
™
488, Alexa Fluor
™
555,
Composite). Layers allow toggling on and off analysis tools.