ThermoFisher Scientific DeviceLink Connect Скачать руководство пользователя страница 29

DeviceLink Connect

 Troubleshooting  |   25

Troubleshooting

This section is a guide to troubleshooting general operational problems.

Symptom

Issue

Potential Cause

Solution

Data 
Transmission

The frequency at 
which data is 
reported from my 
serial cable 
connected 
instrument is 
1 min, but I have 
selected a different 
frequency during 
the setup/
commissioning 
process.

Serial port connected 
instrument.

This is expected. When using a serial connection to 
extract information from your instrument, the data 
transmission is restricted to the logging frequency of 
the information on the instrument itself. The logging 
frequency on instruments is every 1 minute and 
cannot be changed, regardless of selection during 
the commissioning process.

Data 
Transmission

I am missing data 
after a network 
disruption.

Data backfill delay.

When the network connection is interrupted (power 
outage/network loss), the DeviceLink Connect mote 
logs data internally. Once the network connection is 
restored, this logged data is pushed to the cloud. 
Depending on the duration of the network outage 
and subsequently the amount of data needing to be 
backfilled to the cloud, the backfill process may take 
several hours to occur.

Deployment / 
Commissioner

I press “Deploy” 
and the system will 
not complete (go 
Green).

Network ports not 
open.

Ports 80 and 443 must be open on the network for 
inbound/outbound to allow for the communication 
between the DeviceLink Connect and the Thermo 
Fisher Cloud. Configuring the ports open on 
outbound only may still yield an issue as the device 
must receive a confirmation, inbound, from the 
network. 

Network is 
intermittently down.

Return to the Wi-Fi screen to ensure the network is 
active.

No sensor connected.

You must have at least one active sensor connected 
and transmitting data to complete commissioning. 
Ensure at least one sensor is connected.

Other.

Contact technical support.

Factory 
Resetting 
Device

I need to factory 
reset my DeviceLink 
Connect.

Factory Reset 
Procedure via 
Commissioner 
Application.

Follow the 

Getting Started

 section to access your 

Commissioner Application. From the left navigation 
menu in the application, open the “Quality Control” 
section. From this section you will see the perform 
Factory Reset option. Once a Factory Reset is 
conducted, it is recommended to power cycle the 
device.

Содержание DeviceLink Connect

Страница 1: ...332858H01 Revision A February 2022 DeviceLink Connect Configuration and Installation Instruction Guide...

Страница 2: ...amount paid by the client for said product Under no circumstances shall the seller or creator assume responsibility for any dam age or prejudice whatsoever direct or indirect specific or consequentia...

Страница 3: ...endations 4 Applicable Equipment Models 4 DeviceLink Connect Independent Sensors 5 Technical Specifications 6 Mote Specifications 6 Independent Sensor Specifications 7 Getting Started 8 Prerequisites...

Страница 4: ...21 InstrumentConnect Monitoring Application 22 Managing Users 22 Favorites 22 Disconnecting Instruments 23 Scheduling an Instrument 23 Collections 23 Viewing Instrument Data 24 Downloading Data Creati...

Страница 5: ...red as medical device and has therefore not been registered with a medical device regulatory body e g FDA These devices are not to be used for any medical purposes Environmental Operating Conditions T...

Страница 6: ...s use the specific Power Injector adapter provided by Thermo Fisher Scientific Do not open the adapter yourself and do not dismantle internal components or modify them in any manner The adapter does n...

Страница 7: ...ance This device complies with ISED s license exempt RSS standard s Operation is subject to the following two conditions 1 This device may not cause harmful interference and 2 This device must accept...

Страница 8: ...ceLink Connect is not supported For a complete list of compatible equipment by model number go to www thermofisher com DLCequipment Table 1 Equipment Models Equipment Type Brand Equipment Series Name...

Страница 9: ...nk Connect Model Number Additional cable connections required Temperature PT100 DLCPTTEMP001 DLCPTSMRT001 CO2 DLCPTCO2A001 Current 0 50 A DLCPTCURT001 DLCPTSMRT002 Current 0 80 A DLCPTCURT002 DLCPTSMR...

Страница 10: ...recharges in 5 hours or less 2600 mAH 3 6 V Compliant with UL 2054 second edition dated 10 29 2004 including revisions through 2 11 2011 The DeviceLink Connect mote contains a Li Ion battery which IAT...

Страница 11: ...01 n a Temperature RH Temperature RH CO2 Interface USB RS485 RJ45 RS485 RJ45 System Accuracy Temp 1 C RH 3 RH Temp from 10 C to 30 C 0 5 C Temp from 40 C to 10 C 0 6 C RH at temps from 0 C to 40 C an...

Страница 12: ...a supported web browser application and Wi Fi or Ethernet port Installing the DeviceLink Connect Step 1 Setup Your Hardware Connect your DeviceLink Connect mote to the instrument being monitored via t...

Страница 13: ...mote Install any additional independent sensors desired See Installing Independent Sensors section for added instructions Install provided USB Commissioner dongle into any 1 of the 4 USB ports on the...

Страница 14: ...Ethernet Figure 4 LAN port on DeviceLink Connect Mote Turn your DeviceLink Connect mote on by sliding the power switch on the front of the mote to the ON position Figure 5 Mote OFF ON Switch Step 2 Es...

Страница 15: ...proceed directly to the Sensors setup 4 Navigate to the WiFi screen using the left navigation bar or links on the welcome screen a Click the Scan WiFi Networks button to scan for available networks Fi...

Страница 16: ...or freezer incubator etc The asset serial number is used for associate and link data coming from the equipment for display on Thermo Connect c Click Deploy Once the deployment is complete you will see...

Страница 17: ...t 60 seconds and refresh the page Click on the Pending Request banner Figure 14 Pending Request on InstrumentConnect screen 4 Accept the request 3 Figure 15 Accepting request connects to InstrumentCon...

Страница 18: ...er into the rectangular Insert Release slot for the wire that you want to remove Inserting the micro screwdriver into the slot requires some additional downward pressure to actually open the wire clam...

Страница 19: ...ective cover 8 Secure the temperature sensor next to the control sensor In some units use of grommets may be necessary to ensure the sensor tip does not come in contact with exposed metal surfaces ins...

Страница 20: ...r access port 12 Restore power and restart the unit 13 Using the RTD Smart Sensor Cable insert and secure the sensor wires See instructions above 14 Connect the sensor to the desired USB port on the m...

Страница 21: ...oxy Plastic Bonder The mounting surface must be thoroughly cleaned and dried with the included alcohol pad prior to affixing the brackets to the asset Pre requisites Paper towels 70 alcohol solution i...

Страница 22: ...nt the bracket to the asset door Press down firmly with at least 15 psi for 30 seconds to make a good seal a Top of the door is best b All of the adhesive tape must be flush and in contact with the mo...

Страница 23: ...with RADIUS authentication server WPA Enterprise Key management for CCMP and TKIP WPA and full IEEE 802 11i RSN WPA2 RSN PMKSA caching pre authentication IEEE 802 11r EEE 802 11w Wi Fi Protected Setu...

Страница 24: ...one NTP server in order to maintain the correct time The default NTP server URLs built in to the mote may work or the customer IT group may provide their own NTP server URL s 1 Select NTP from the men...

Страница 25: ...mote is connecting via an Ethernet connection then uncheck the box directly under PoE Adapter Configuration 3 When either box is unchecked a form will appear with several blank fields 4 The customer I...

Страница 26: ...From this dashboard you are able to see all the instruments connected to your Thermo Connect account Figure 26 InstrumentConnect dashboard Managing Users The first user to connect to an instrument is...

Страница 27: ...Scheduling an Instrument Collections From the top menu go to the Collections tab This tool allows an Admin to group instruments and share them with other users on Thermo Connect Creating a New Collec...

Страница 28: ...at the top of the graph event log container Indicate which report type is desired CSV or PDF Figure 33 Create Report You will see a notice at the top of the screen indicating your report generation is...

Страница 29: ...ork outage and subsequently the amount of data needing to be backfilled to the cloud the backfill process may take several hours to occur Deployment Commissioner I press Deploy and the system will not...

Страница 30: ...he notification in InstrumentConnect but I am not seeing the new device Page has not been refreshed Refresh the InstrumentConnect web browser page and or application window If after a refresh your dev...

Страница 31: ...ioner should not need to be modified during normal setup unless directed by Thermo Fisher Scientific If modifications have been made attempt a factory reset of the mote see Factory Reset instructions...

Страница 32: ...nt left of the mote to the On position The switch is slightly recessed Loss of AC power at the outlet Consult facilities to ensure AC power is provided to the device Power DeviceLinkConnect not operat...

Страница 33: ...on the mote Sensors not assigned properly in Commissioner setup Reinsert the USB commissioning dongle this may require you to temporarily remove a sensor if all 4x USB ports are in use Follow the step...

Страница 34: ...t application to Accept your new device connection If you do not receive this notification within 30 minutes re commission the device following the steps in the Commissioner application per the Gettin...

Страница 35: ...374 91 22 6716 2200 India 91 22 6716 2200 91 22 6716 2200 Italy 32 02 95059 552 39 02 95059 552 432 254 375 Japan 81 3 5826 1616 81 3 3816 3355 Netherlands 31 76 579 55 55 31 76 571 4440 Nordic Baltic...

Отзывы: