Orion Benchtop pH/ISE Meter Instruction Manual
Troubleshooting
71
Error
Code
Problem
Probable Remedy
E-22
Calibration
The millivolts being measured are
Std Error
the same for two different standards.
Check that two different standards
or buffers are being used and that
the correct one is being measured.
1. Make fresh standards or buffers
and repeat the last calibration
point or the entire calibration.
E-23
Bad Slope
pH electrode slope not in the range
of 80% to 120%.
1. Press any key to acknowledge
and repeat calibration using
fresh buffers.
2. Clean electrode(s) and refill
reference electrode(s).
3. Refer to electrode instruction
manual for instructions on how
to check electrode operation.
E-26
Datalog Full
The datalogger is full and cannot
store any more points. A maximum
of 25 points may be stored.
1. Transfer data to a printer
or computer.
2. Clear data from memory
.
For Orion 710Aplus:
a. Go to 6-3 in the setup menu and
transfer data to the printer or
computer by pressing
then
yes.
b. Go to 6-4 in the setup menu and
press then
yes.
Once the
data has been cleared additional
points may be logged.
c. For Orion 520Aplus, 525Aplus,
720Aplus, and 920Aplus: Press
2nd
then
options
to access the
options menu. Follow the
prompts to transfer and clear the
datalogger.
▼
▼
▼
▼
Содержание Orion 410Aplus
Страница 6: ...2 Orion Benchtop pH ISE Meter Instruction Manual...
Страница 84: ...80 Orion Benchtop pH ISE Meter Instruction Manual Certificate of Conformity...
Страница 92: ...Orion Benchtop pH ISE Meter Instruction Manual 88...
Страница 104: ...Orion Benchtop pH ISE Meter Instruction Manual 100...
Страница 118: ...Orion Benchtop pH ISE Meter Instruction Manual Overview of Meter Functions 520Aplus 525Aplus 720Aplus and 920Aplus 114...
Страница 120: ...Orion Benchtop pH ISE Meter Instruction Manual 116...
Страница 122: ...Orion Benchtop pH ISE Meter Instruction Manual 118...
Страница 126: ...Orion Benchtop pH ISE Meter Instruction Manual 122...