![Thermo Orion 410Aplus Скачать руководство пользователя страница 24](http://html1.mh-extra.com/html/thermo/orion-410aplus/orion-410aplus_instruction-manual_1104228024.webp)
Orion Benchtop pH/ISE Meter Instruction Manual
Calibration and Measurement of pH
20
pH Calibration Procedure
1. Connect electrode(s) to meter. Choose buffers that will bracket the
expected sample pH.
2. Press
mode
until the pH mode indicator is displayed.
3. Rinse electrode(s) and place into buffer.
4. Press
2nd
then
cal
to begin calibration. The date and time of the last
calibration will be displayed.
5. When READY is displayed next to the reading, indicating electrode
stability, the reading will flash. Press
yes
. For manual calibration
use or to select the correct value, then press
yes
to accept each
digit. After accepting each digit, press
yes
store buffer value. The
meter display freezes for three seconds. The meter automatically
switches to buffer two, indicated by the P2 on the display.
6. Repeat steps 3 through 5 for each buffer.
7. After entering the final buffer value, press
measure
. The electrode
slope will be displayed. SLP appears in the lower field while the
actual electrode slope, in percent, appears in the main field. After the
third buffer point, the meter automatically displays the calibration
slope and advances to the measure mode. MEASURE is displayed
above the main field.
8. Rinse electrode(s) and place into sample. Record pH directly from the
main meter display. Temperature is displayed in the lower field
when the READY prompt displays.
▼
▼
Содержание Orion 410Aplus
Страница 6: ...2 Orion Benchtop pH ISE Meter Instruction Manual...
Страница 84: ...80 Orion Benchtop pH ISE Meter Instruction Manual Certificate of Conformity...
Страница 92: ...Orion Benchtop pH ISE Meter Instruction Manual 88...
Страница 104: ...Orion Benchtop pH ISE Meter Instruction Manual 100...
Страница 118: ...Orion Benchtop pH ISE Meter Instruction Manual Overview of Meter Functions 520Aplus 525Aplus 720Aplus and 920Aplus 114...
Страница 120: ...Orion Benchtop pH ISE Meter Instruction Manual 116...
Страница 122: ...Orion Benchtop pH ISE Meter Instruction Manual 118...
Страница 126: ...Orion Benchtop pH ISE Meter Instruction Manual 122...