Orion Benchtop pH/ISE Meter Instruction Manual
Overview of Meter Functions 520Aplus,525Aplus,720Aplus and 920Aplus
108
Key Decimal Point Data/print modes
Press the
decimal point
to set the datalog function and the print mode. See
the descriptions below for instructions. Print mode is presented first, then
the datalog function.
Printing
Press
data/print modes
to select a print option, then follow the prompts:
Prompt
Keypress/Action
Press
data/print modes
.
1. MANUAL PRINT
The current print mode is displayed. Wait for
the prompt.
2. SET PRINT MODE?
Press
yes
.
3. 1-RDY
The three print options are displayed
2-TIM
along with current setting:
3-MAN
1 For Print on ready.
2 For Print at a timed interval.
3 For manual Print.
Enter
1
,
2
, or
3
and press
yes
.
4. MANUAL PRINT
Your choice is verified, then the meter
proceeds to steps 6-10 if timed print selected.
5. PRT INTRVL HRS
Enter the print interval hours setting, 00 to
23, then press
yes
.
6. PRT INTRVL MIN
Enter the print interval minutes setting, 00 to
59, then press
yes.
7. PRT INTRVL SEC
Enter the print interval seconds setting, 00 to
59, then press
yes
.
8. PRINT INTERVAL
The print interval is verified then the meter
continues.
9. SET LOG MODE?
Press
yes
to continue (see below) or
no
to exit.
Содержание Orion 410Aplus
Страница 6: ...2 Orion Benchtop pH ISE Meter Instruction Manual...
Страница 84: ...80 Orion Benchtop pH ISE Meter Instruction Manual Certificate of Conformity...
Страница 92: ...Orion Benchtop pH ISE Meter Instruction Manual 88...
Страница 104: ...Orion Benchtop pH ISE Meter Instruction Manual 100...
Страница 118: ...Orion Benchtop pH ISE Meter Instruction Manual Overview of Meter Functions 520Aplus 525Aplus 720Aplus and 920Aplus 114...
Страница 120: ...Orion Benchtop pH ISE Meter Instruction Manual 116...
Страница 122: ...Orion Benchtop pH ISE Meter Instruction Manual 118...
Страница 126: ...Orion Benchtop pH ISE Meter Instruction Manual 122...