Orion Benchtop pH/ISE Meter Instruction Manual
Overview of Meter Functions-420Aplus, 710Aplus
97
Code
Function
Description
3-4
SET DATE
Sets the current date. When the code 3-4 is
displayed in the lower field, the current date
(month: day) is displayed in the main field.
Press
yes
, or change the date as needed, then
press
yes
. Next the current year is displayed.
Press
yes
or change as required, then
press
yes
.
4-1
HIGH
Sets the high alarm limit for the current
ALARM
measurement mode. When the reading
LIMIT
exceeds the set limit the alarm will
(710Aplus only)
ring. The maximum limits are 19.999 for
pH, 19900 for concentration, +1600.0 for
mV, and +1999.9 for relative millivolts.
Press or to change, then press
yes
.
4-2
LOW
Sets the low alarm limit for the current
ALARM
measurement mode. When the reading falls
LIMIT
below the set limit the alarm will ring.
(710Aplus only)
The minimum limits are -2.000 pH, 0.0
for Concentration, -1600.0 for mV and
-1999.9 for relative millivolts. Press or
to change, then press
yes
.
5-1
Sets the print mode. When 5-1 is displayed
MODE
in the lower field the current print mode is
displayed in the main field. The options are:
1:
manual print, indicates no automatic
output to the printer and the user may
print on command by pressing
2nd
then
print.
2:
Print on Ready, the meter will send
information to the printer whenever
the electrode signal reaches stability.
3:
Print on a timed interval, printing
occurs at a preset timed interval. Press
or to change, then press
yes.
The default setting is 1: manual print.
▼
▼
▼
▼
▼
▼
Содержание Orion 410Aplus
Страница 6: ...2 Orion Benchtop pH ISE Meter Instruction Manual...
Страница 84: ...80 Orion Benchtop pH ISE Meter Instruction Manual Certificate of Conformity...
Страница 92: ...Orion Benchtop pH ISE Meter Instruction Manual 88...
Страница 104: ...Orion Benchtop pH ISE Meter Instruction Manual 100...
Страница 118: ...Orion Benchtop pH ISE Meter Instruction Manual Overview of Meter Functions 520Aplus 525Aplus 720Aplus and 920Aplus 114...
Страница 120: ...Orion Benchtop pH ISE Meter Instruction Manual 116...
Страница 122: ...Orion Benchtop pH ISE Meter Instruction Manual 118...
Страница 126: ...Orion Benchtop pH ISE Meter Instruction Manual 122...