6
•
Operation
Variable Wavelength Detectors (VC-D40, VF-D40)
Operating Manual
Page 89
Parameters
Description
Spectra scanning
To record the absorbance spectra of substances that are
used in an analysis, for example, to detect peak
impurities, or to identify optimum wavelengths for
method development, perform spectra scanning:
ScanStartWavelength
sets the wavelength at which the
scan should start.
ScanEndWavelength
sets the wavelength at which the
scan should end.
ScanSpeed
sets the speed at which the wavelength range
is scanned.
ScanBaseline
saves a baseline scan based on the current
settings.
Scan
saves an absorbance spectrum based on the current
settings.
Note:
Before you can record an absorbance spectrum,
perform a
ScanBasline
with the same wavelength settings
that you want to use for recording the absorbance
spectrum.
For further information about the parameters and
commands, refer to the
Chromeleon Help
.
UV lamp
VIS lamp
Turn on the UV and/or VIS lamp before starting an
analysis (
UV_Lamp = On, Visible_Lamp =On
). This setting
allows turning on and off the UV and VIS lamp of the
detector.
To turn the UV lamp on automatically after power on, set
AutoactivateUV_Lamp
to
On
.
To turn the VIS lamp on automatically after power on, set
AutoactivateVisible_Lamp
to
On
.
UV_VIS
The
UV_VIS
signal channels (
UV_VIS_x
) that are available
for data acquisition are listed in the dialog box for the
detector in the Instrument Configuration Manager. Select
the required signal channels.
Recording at more than one signal channel, thus at more
than one wavelength, sets the detector into multi-channel
mode. For details, see
Wavelength
Sets the wavelength in nm. Select the wavelength at the
absorption maximum of the sample component to be
analyzed.
Wavelength calibration
If the wavelength validation failed, perform a calibration
of wavelengths.
The 0-order radiation, which corresponds to 000 nm of
the wavelength scale, is used to calibrate the wavelength.
For further information, see
.
Wavelength validation
Perform the wavelength validation as required. A
holmium-oxide glass filter is moved into the light path of
the lamp and is used to validate the wavelength accuracy.
For further information, see
.
Содержание Thermo Scientific Vanquish VC-D40
Страница 4: ...Contacting Us Page 4 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Страница 10: ...Contents Page 10 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Страница 98: ...6 Operation Page 98 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Страница 136: ...7 Maintenance and Service Page 136 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Страница 148: ...8 Troubleshooting Page 148 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Страница 154: ...9 Specifications Page 154 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Страница 168: ...11 Appendix Page 168 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Страница 173: ...Index Variable Wavelength Detectors VC D40 VF D40 Operating Manual Page 173 ...
Страница 174: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...