Operation
Variable Wavelength Detectors (VC-D40, VF-D40)
Operating Manual
Page 87
Consider the following, when operating the detector in multi-
wavelength mode:
•
Each additional channel reduces the maximum possible data
collection rate, and leads to an increased baseline noise.
•
Switching the optical grating permanently between the wavelengths
can lead to wear on the mechanical parts of the detector in the long
term. The multi-channel operation mode is therefore
only
recommended for method development and peak-purity analysis,
but
not
for continuous routine operation.
•
As an alternative, you can switch the wavelengths as often as
required in a single data channel in the Chromeleon software.
For further information on wavelength switching and single-
wavelength operation, see
the
Chromeleon Help
.
TIP
The optical filters for eliminating higher-order wavelengths are
moved automatically into the light path. This applies for multi-channel
mode, too.
Содержание Thermo Scientific Vanquish VC-D40
Страница 4: ...Contacting Us Page 4 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Страница 10: ...Contents Page 10 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Страница 98: ...6 Operation Page 98 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Страница 136: ...7 Maintenance and Service Page 136 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Страница 148: ...8 Troubleshooting Page 148 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Страница 154: ...9 Specifications Page 154 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Страница 168: ...11 Appendix Page 168 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Страница 173: ...Index Variable Wavelength Detectors VC D40 VF D40 Operating Manual Page 173 ...
Страница 174: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...