1
Introduction and Description
Technical Data
Thermo Scientific
Sorvall LYNX 4000 / 6000
3
Technical Data
The technical data of the Sorvall LYNX 4000 / 6000 centrifuge is listed in the following table.
Table 1. Technical Data
Feature
Sorvall LYNX 4000
Sorvall LYNX 6000
Environmental Conditions
-
Use in interior Spaces
-
Altitudes of up to 2,000 m above Sea Level
-
Max. relative humidity 85
% up to 30 °C
Permissible Ambient Temperature
+5 °C to +35 °C
+5 °C to +35 °C
Overvoltage Category
II
II
Pollution Degree
2
2
Heat Dissipation /h
*
2.0 kWh / 6,830 Btu / 7,200 kJ
1.2 kWh / 4,100 Btu / 4,320 kJ
IP (statement of the degree of protection
according to IEC 60529)
20
20
Run Time
99 hours:59min, hold
99 hours:59min, hold
Maximum Speed n
max
24000 rpm (depending on
the rotor)
29000 rpm (depending on
the rotor)
Minimum Speed n
min
500 rpm
500 rpm
Maximum RCF-Value at n
max
68905 x g
100605 x g
Maximum kinetic Energy
< 203 kJ
< 203 kJ
Noise Level for swinging Bucket at
maximum Speed
< 61 dB (A)
**
*Rotor A27-8x50, Typical Run 4 °C at 24,000 rpm 30 minutes, 60 Hz +5
%, 30 min stand by 4° C (lid closed)
**Measures at 1 m distance and at 1.6 m height
< 61 dB (A)
**
Noise Level for Fixed Angle Rotors at
maximum Speed
< 59 dB (A)
**
< 59 dB (A)
**
Temperature Setting Range
-10 °C to +40 °C
-20 °C to +40 °C
Dimensions
Height with Door closed (incl. GUI)
Height with Door open
Width
Depth
1048 mm
1531 mm
700 mm
805 mm
1048 mm
1531 mm
700 mm
805 mm
Weight without Rotor
28
5
kg
2
95
kg
Environmental Conditions
during Storage and Shipping
Temperature:
-10 °C to +55 °C
Humidity: 15
% to 85
%
Temperature:
-10 °C to +55 °C
Humidity: 15
% to 85
%
Содержание Sorvall LYNX 4000
Страница 18: ...1 Introduction and Description 8 Sorvall LYNX 4000 6000 Thermo Scientific ...
Страница 28: ...2 Before Use 18 Sorvall LYNX 4000 6000 Thermo Scientific ...
Страница 34: ...3 Control Panel 24 Sorvall LYNX 4000 6000 Thermo Scientific ...
Страница 71: ...Thermo Scientific Sorvall LYNX 4000 6000 61 Index ...