1
Introduction and Description
Mains Supply
6
Sorvall LYNX 4000 / 6000
Thermo Scientific
Mains Supply
The following table contains an overview on the electrical contact data of the
Sorvall LYNX 4000 / 6000 centrifuge. This data is to be taken into consideration when selecting the
mains connection socket.
1
For the 200-240 V units the inrush current is 120 A for up to 1 second during the refrigeration start period. For the 380 / 400 V units it is 60 A. Circuit breakers,
whether thermal or magnetic actuated, must have a delay type applicable for the start of motors.
2
Use a 25 A or 32 A Trip Char. C circuit breaker (D or K are also applicable).
3
For 380 / 400 V, 3-phase (unbalanced load, no neutral used) use a 16 A Trip Char. C circuit breaker (D or K are also applicable).
4
For North America: use for example GES-9888 30 A.
Table 4. Electrical Connection Data
Mains Voltage
in V
Frequency
in Hz
Rated Current
in A
Power Consump-
tion in W
Building Fuse in
A
1
Equipment
Fuse in A
200
50
24
4
6
00
30
30
208
50
23
4
60
0
30
30
220
50
21
4600
32
2
30
230
50
2
0
4600
32
2
30
240
50
19
4600
32
2
30
380
50
13.5
4600
16
3
16
400
50
12.5
4600
16
3
16
200
60
24
48
0
0
30
30
208
60
2
3
48
00
30
30
220
60
2
1
4800
3
2
2
30
230
60
2
0
4800
3
2
2
30
240
60
19
4800
3
2
2
30
380
60
13.5
4800
16
3
16
400
60
12.5
4800
16
3
16
415
5
0
11.5
4
6
00
16
3
16
415
60
11.5
4800
16
16
Содержание Sorvall LYNX 4000
Страница 18: ...1 Introduction and Description 8 Sorvall LYNX 4000 6000 Thermo Scientific ...
Страница 28: ...2 Before Use 18 Sorvall LYNX 4000 6000 Thermo Scientific ...
Страница 34: ...3 Control Panel 24 Sorvall LYNX 4000 6000 Thermo Scientific ...
Страница 71: ...Thermo Scientific Sorvall LYNX 4000 6000 61 Index ...