5
Maintenance and Care
Cleaning the Touchscreen
38
Sorvall LYNX 4000 / 6000
Thermo Scientific
Cleaning the Touchscreen
1. Pull out the power supply plug.
2. Clean the touchscreen using a dry microfiber cloth.
3. If necessary moisten the microfiber cloth and wipe the touchscreen again.
Cleaning the Condenser Filter
The Sorvall LYNX 4000 / 6000 centrifuge has two filter mats to prevent dust from entering the
centrifuge. In order to clean them proceed as follows:
1. Loosen the four screws of both venting grids on the right side of the centrifuge.
2. Remove the venting grids and the filter mats (75000014).
3. Use a vacuum cleaner on both sides of both filter mats.
4. Inserting the filter mats again.
5. Reinstall the venting grids.
Figure 17. Venting Grids
CAUTION
The fins of the condenser are very sharp. Make sure to wear protective gloves
when removing the filter mats.
Screws
Filter Mats