32
TissueWave 2 Operator Guide
The specific service requirements are as follows:
120V Model - 15A Service Minimum (20A Recommended)
230V Model - 10A Service Minimum (15A Recommended)
The Shandon TissueWave
Vent System
is designed to connect to a
positively vented (powered) laboratory hood
or
building vent
stack
.
A
4” (100 mm) Diameter Vent Hose
is included with the unit for this
purpose.
Caution should be used regarding the length and number of bends in
this piece of ducting.
In general, right angle bends should be avoided and the total length
of the hose should be kept to
less than 8 feet (2.4 m)
.
The Shandon TissueWave draws cooling air through a
Filter
located
on the right-hand, front, bottom surface of the unit and exhausts it
through a port on the back wall.
Care should be taken to keep these air vents from being obstructed.
The protrusion of the
Vent System Housing
on the back will usually
provide adequate clearance for the exhaust outlet.
Periodic cleaning of the
Inlet Screen
will also help ensure sufficient
cooling for the
Magnetron
and other heat sensitive devices in the
unit.
If airflow is restricted, the unit will not operate properly and
the life of the electrical parts will be shortened.
•
•
Ventilation
Requirements
Ventilation
Requirements
Cooling
Requirements
Cooling
Requirements
!!