23
TissueWave 2 Operator Guide
The
Shandon TissueWave 2 Vacuum Chamber
is designed to
accept
Shandon Excelsior™ Processing Baskets
, which allow up
to
74 Tissue Cassettes
to be processed at once.
The
Processing Basket
comprises of two sections, each capable of
containing
37 Tissue Cassettes
in an ordered manner.
The procedure for filling and loading the
Processing Baskets
is as
follows:
Arrange the
Tissue Cassettes
as required in the
Processing
Basket
halves.
Place the two halves together as shown, and insert into the
Vacuum Chamber
Base
.
Fill to the required level with
Processing Fluid
.
Insufficient fluid levels may cause the probe to
overheat and prevent the unit from operating.
If vacuum is required, place the
Vacuum Chamber Lid
on
the Vacuum Chamber Base.
Ensure the lid forms a tight seal around the rim of the
base, as poor sealing will negate the vacuum effects.
It is recommended that the
Vacuum Chamber
be placed
onto a
Microwavable Plastic Tray
to facilitate safe
removal after processing.
•
•
•
•
•
Loading the
Shandon
TissueWave 2
Vacuum Chamber
Loading the
Shandon
TissueWave 2
Vacuum Chamber
!!
!!