4 - Programming
There are some concepts about the Shandon TissueWave 2 unit
which should be understood prior to beginning programming.
The Shandon TissueWave 2 has two
Control Modes
which can be
used:
Temperature Control
Power Control
Use the
Function Keys
to select
Temperature (T)
or
Power (P)
when prompted.
In
Temperature Control Mode
, the
Output Power
is continually
adjusted to maintain a particular
Temperature Profile
.
The
Temperature Profile
can be further defined using the
Timer
Mode
field.
The
Timer Mode
options are:
Time @
- Used to define the
Procedure Time
from the time
when the
Solution
reaches the required temperature.
Time T
- Used to define the
Total Procedure Time
including
the
Ramp Time
required to achieve the correct temperature.
In
Power Control Mode
, the
Output Power
is held at a
User-
Defined Preset Value
.
Solution Temperature
is ignored as a control factor.
When using most
Published Procedures
, the user will need to use
the
Temperature Control Mode
.
•
•
•
•
Control
Parameters
Control
Parameters